FDA Medical Device Recalls (openFDA)
C.R. Bard Inc: HydroSil Go Intermittent Urinary Catheters, Catalog Numbers: a) REF 71808, b) REF 71810, c) REF 71812, d) REF 71814, e) REF 71816, f) REF 71818
- Recall number
- Z-1701-2023
- Recalling firm
- C.R. Bard Inc
- Product
- HydroSil Go Intermittent Urinary Catheters, Catalog Numbers: a) REF 71808, b) REF 71810, c) REF 71812, d) REF 71814, e) REF 71816, f) REF 71818
- Status
- Open, Classified
- Initiated
- 2023-04-20
- Posted
- 2023-06-08
- Terminated
- not on file
- Reason
- Some catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.
- Root cause
- Under Investigation by firm
- Action
- The firm issued an URGENT MEDICAL DEVICE RECALL to its consignee beginning on 04/20/2023 by Fed Ex and email. The notice explained the problem, risk, and requested the following actions be taken: Discontinue distribution of the product and dispose of all affected product, Distributors are directed to notify their customers. For further assistance please contact: 1-844-8BD- LIFE (1-844-823-5433) Say "Recall" when prompted; M-F 8am - 5pm CT
- Quantity
- 308520 devices
- Distribution
- Worldwide
- Lot, serial or model codes
- a) REF 71808, Lot Numbers: JUGU9026; b) REF 71810, Lot Numbers: JUGT9154, JUGU9053, JUGV9130, JUGV9133; c) REF 71812, Lot Numbers: JUGT0236, JUGT1689, JUGT2702, JUGU1955, JUGU1968, JUGV9158; d) REF 71814, Lot Numbers: JUGT3326; e) REF 71816, Lot Numbers: JUGU9023; f) REF 71818, Lot Numbers: JUGT0288, JUGU9022
- 510(k) numbers
- ["K150345"]
- PMA numbers
- not on file
- Product code
- EZD
- Firm FEI
- 1018233
- Firm city
- Covington
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as