FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.
- Recall number
- Z-1675-2023
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.
- Status
- Open, Classified
- Initiated
- 2023-04-06
- Posted
- 2023-06-02
- Terminated
- not on file
- Reason
- There is a low possibility that an electrical connector in the MR gradient coil will overheat on Elekta Unity systems.
- Root cause
- Process design
- Action
- The recalling firm issued the Field Safety Notice dated March 2023 via email on 4/6/2023 describing the problem, the clinical impact, and the recommended user action, which is that no actions are required to be taken by the user. The Unity system is safe for clinical use without restriction. The consignee is informed the letter is to notify them there will be action taken on the Elekta Unity system to further reduce the potential risk of overheating within the MR Gradient Coil which could lead to system downtime. The letter is to be posted in a place accessible to all users and appropriate personnel are to be advised on the content of the letter. The letter contains Elekta Corrective Actions, which says the update will be provided via Elekta's Field Change Order process once available and that their local Elekta service representative will contact the consignee to arrange for this activity to be completed. An acknowledgement Form was enclosed to be returned to Elekta immediately upon receipt but no later than within 30 days. The device serial number and location or site of the device is to be reported and the consignee is to sign it to acknowledge they have read and understand the notice and accept the implementation of any given recommendation. UPDATED 12-JUN-23 An updated Important Field Safety Notification was released to customers by email in which supersedes the previous Notice under ref: 200-01-801-012 released on 06-APR-23. The new FSN includes additional information under Recommended User Action.
- Quantity
- 18 devices
- Distribution
- US Nationwide distribution in the states of FL, IA, NJ, NY, OK, PA, TN, WI, MI, LA, KS, NM and TX.
- Lot, serial or model codes
- UDI GTIN 05060191071598, serial numbers: 14117-007/600034, 10632-010/600036, 10817-011/600010, 10905-008/600037, 11198-010/600047, 11198-011/600048, 11830-010/600064, 11888-004/600072, 12108-007/600019, 13371-002/600021, 13553-006/600003, 30003193-003/600074, 13120-017/600073, 11328-019/600059, 14232-001/600095, 12269-004/600100, 11008-003/600104, 13212-007/600106
- 510(k) numbers
- K212114
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28