Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: BASE, S-LIFT Instrument Case Product Usage: System cases are designed to hold and secure all instrumentation to be used with a surgical system. Case and caddy components are used to contain implants intended for intervertebral body fusion and fixation of the spine.

Recall number
Z-1664-2013
Recalling firm
SpineFrontier, Inc.
Product
BASE, S-LIFT Instrument Case Product Usage: System cases are designed to hold and secure all instrumentation to be used with a surgical system. Case and caddy components are used to contain implants intended for intervertebral body fusion and fixation of the spine.
Status
Terminated
Initiated
2013-05-17
Posted
2013-07-05
Terminated
2016-05-17
Reason
The S-LIFT Instrument Case contained a bracket with peeling nylon coating.
Root cause
Component design/selection
Action
Spine Frontier sent an Urgent Advisory Notice letter dated On June 7, 2013 to all consignees via email. The letter identified the affected product, problem and actions to be taken. The letter informed customers that a modification has been implemented for the improved durability of S-LIFT Case Brackets and that SpineFrontier has initiated removal of these cases from the field. An acknowledgement form is to be signed and returned from each recipient of the Advisory Notice. Upon receipt of the acknowledgement form and RMA request for part return, replacement parts will be shipped. For questions call 978-232-3990.
Quantity
12
Distribution
USA Nationwide including the states of: FL, CO, Wash DC, VA, KS, TX, OH, TN, OK, CA, IL, MO, and MN.
Lot, serial or model codes
Lot/Serial Numbers: 7201.007 and 7201.007-1
510(k) numbers
["K092815"]
PMA numbers
not on file
Product code
FSM
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register