Skip to the content

FDA Medical Device Recalls (openFDA)

Neusoft Medical Systems Co., Ltd.: NeuViz 64 Multi-Slice CT Scanner System (consist if two variants: NeuViz 64e, NeuViz 64i)

Recall number
Z-1650-2016
Recalling firm
Neusoft Medical Systems Co., Ltd.
Product
NeuViz 64 Multi-Slice CT Scanner System (consist if two variants: NeuViz 64e, NeuViz 64i)
Status
Terminated
Initiated
2015-05-02
Posted
not on file
Terminated
2016-11-04
Reason
It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn't meet the patient position description, when the scanner position was set to "Right" in the system setting - scanner options.
Root cause
Device Design
Action
An upgraded software version (1.0.5.2619+P11) was released on May 7, 2015. The FCO was released to the service engineers of Neusoft Medical Systems USA, which is subsidiary of the reporting firm, and the engineers performed FCO for each affected device together with the distributors' engineers. The corrections (FCOs to upgrade software) on all the affected systems in the USA have been completed. At present (4/27/16) there is no longer this issue on the systems, so no further actions are needed for customers/distributors. However, the recalling firm shall inform the direct accounts of this fix before May 5, 2016. The recalling firm will submit the communication evidence ASAP once they have evidence of the communication. For further questions, please call (281) 453-1205.
Quantity
7 units
Distribution
US Distribution - Including Puerto Rico and the states of IL, CT, SC, NE
Lot, serial or model codes
Serial Numbers: N64E120022E, N64E130039E, N64I130051E, N64E130062E, N64E130063E, N64E130066E, N64I140077E
510(k) numbers
["K121792"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3003575350
Firm city
Shenyang
Firm state
not on file
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register