Skip to the content

FDA Medical Device Recalls (openFDA)

Cytrellis Biosystems, Inc.: Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4

Recall number
Z-1648-2023
Recalling firm
Cytrellis Biosystems, Inc.
Product
Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4
Status
Completed
Initiated
2023-04-07
Posted
2023-05-26
Terminated
not on file
Reason
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.
Root cause
Use error
Action
Cytrellis issued Medical Device Update / Product Notification Letter via email on 4/7/23. Letter states reason for recall, heath risk and action to take: Important information is included in the new Quick Reference Guide, with emphasis on the following points: "DO NOT impede tubing on handpiece. "DO NOT apply forceful downward pressure. "DO NOT TOUCH needle cartridge tubing while treating patient. Your PDM will be reaching out soon to set up time with you and your staff to review the Quick Reference Guide, and we would also like to invite you to join our medical team in one of two live informative webinars to review best practices and answer any questions that you might have. These webinars will be hosted by our Medical Director, Dr. Ashish Bhatia, along with the Cytrellis Medical Affairs team. Invitations to the webinars will be sent shortly. For additional information or inquiries, please email medicalinformation@cytrellis.com
Quantity
89 units
Distribution
Nationwide
Lot, serial or model codes
UDI-DI: B933ELCON010 All serial numbers
510(k) numbers
["K202517"]
PMA numbers
not on file
Product code
QAI
Firm FEI
3020704367
Firm city
Woburn
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register