Smiths Medical ASD, Inc.: Portex Low Dead Space Connector with Sideport, 3.5mm Product Usage: The connector is used to attach the endotracheal tube to the patient breathing circuit. The low dead space connector minimizes mechanical dead space and subsequent carbon dioxide rebreathing. The connector's sideport permits access for administration of artificial surfactant, airway pressure monitoring, gas sampling, and instillation of sterile saline for airway leakage.

Recall number
Z-1641-2014
Recalling firm
Smiths Medical ASD, Inc.
Product
Portex Low Dead Space Connector with Sideport, 3.5mm Product Usage: The connector is used to attach the endotracheal tube to the patient breathing circuit. The low dead space connector minimizes mechanical dead space and subsequent carbon dioxide rebreathing. The connector's sideport permits access for administration of artificial surfactant, airway pressure monitoring, gas sampling, and instillation of sterile saline for airway leakage.
Status
Terminated
Initiated
2014-04-10
Posted
2014-06-11
Terminated
2016-11-29
Reason
One lot of 3.0mm sized Connectors, Product Reorder No. 10035-05PS, Lot No. 2553426 were placed in packages labeled as size 3.5mm.
Root cause
Packaging process control
Action
Smiths Medical sent an Urgent Medical Device Recall Notice dated April 10, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory and quarantine all affected products and to complete the attached Response Form and return it to Smith Medical within 7 days of receipt of the notice. Customers with questions were instructed to contact Smiths Medical's Customer Service Department at +1-800-258-5361.
Quantity
100 devices
Distribution
US distribution in the state of North Carolina.
Lot, serial or model codes
Lot or Serial No: 2553426
510(k) numbers
K790312
PMA numbers
not on file
Product code
BTR
Firm FEI
1217052
Firm city
Keene
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source