Beckman Coulter Inc.: GLU Reagent (GLU cartridge assay) used on Synchron LX and UniCel DxC Synchron Clinical Systems, Part Number 442640; SYNCHRON and UniCel Systems and Part Numbers: Synchron LX20, Part Number: 466200, Synchron LX20 PRO, Part Number: 476100, Synchron LXI 725, Part Number: 476501, UniCel DxC 600, Part Number: A10405, UniCel 600 PRO, Part Number: A11810, UniCel DxC 600i, Part Number: A27318, UniCel DxC 800, Part Number: A11816, UniCel DxC 800 PRO, Part Number: A11812, UniCel DxC 880i, Part Number: A59102, UniCel DxC 660i, Part Number: A64871, UniCel DxC 680i, Part Number: A64903, UniCel DxC860i, Part Number: A64935 (1) The Synchron LX and UniCel DxC Synchron Clinical Systems are fully automated, computer controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (Sample type is chemistry dependent). (2) GLUCm reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1 and 2, is intended for the quantitative determination of Glucose concentration in human serum, plasma, urine or cerebrospinal fluid (CSF).

Recall number
Z-1619-2011
Recalling firm
Beckman Coulter Inc.
Product
GLU Reagent (GLU cartridge assay) used on Synchron LX and UniCel DxC Synchron Clinical Systems, Part Number 442640; SYNCHRON and UniCel Systems and Part Numbers: Synchron LX20, Part Number: 466200, Synchron LX20 PRO, Part Number: 476100, Synchron LXI 725, Part Number: 476501, UniCel DxC 600, Part Number: A10405, UniCel 600 PRO, Part Number: A11810, UniCel DxC 600i, Part Number: A27318, UniCel DxC 800, Part Number: A11816, UniCel DxC 800 PRO, Part Number: A11812, UniCel DxC 880i, Part Number: A59102, UniCel DxC 660i, Part Number: A64871, UniCel DxC 680i, Part Number: A64903, UniCel DxC860i, Part Number: A64935 (1) The Synchron LX and UniCel DxC Synchron Clinical Systems are fully automated, computer controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (Sample type is chemistry dependent). (2) GLUCm reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1 and 2, is intended for the quantitative determination of Glucose concentration in human serum, plasma, urine or cerebrospinal fluid (CSF).
Status
Terminated
Initiated
2010-08-10
Posted
2011-03-09
Terminated
2013-10-17
Reason
The recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be hemolysed to a degree sufficient to cause interference with glucose estimation and are a special at-risk population, (2) Other hemolysed samples may be similar
Root cause
Other
Action
The firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated August 10, 2010 with an enclosed fax-back Customer Response Form to the customers. The letter provided the customers with an explanation of the problem identified, the impact and an action to be taken. The customers were instructed to run samples using another method/technology and to complete and return the enclosed FAX BACK RESPONSE FORM via fax to (714) 961-4232 or mail to: Beckman Coulter, Inc., Attn: Regulatory Affairs, 250 S. Kraemer Blvd., PO Box 8000, Brea, CA 92822. Firms with questions regarding this Product Corrective Action letter, were instructed to call our Customer Support Center (Hotline) at (800) 854-3633 in the United States and Canada http://www.beckmancoulter.com/customersupport/support/default.asp.
Quantity
81,871 units
Distribution
Worldwide distribution: USA and country: Canada.
Lot, serial or model codes
All non-expired lots
510(k) numbers
["K011213","K023049","K042291","K060256","K883181","K965240"]
PMA numbers
not on file
Product code
CEM
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register