Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Spine: Stryker Mantis Rod Inserter Inner Shaft; Non Sterile; Manufactured by Stryker Spine SAS, Cestas France. Distributed by Stryker Spine, Allendale, NJ 07401-1677. The MANTIS Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: degenerative disk disease, spodylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis and failed previous fusion.

Recall number
Z-1612-2009
Recalling firm
Stryker Spine
Product
Stryker Mantis Rod Inserter Inner Shaft; Non Sterile; Manufactured by Stryker Spine SAS, Cestas France. Distributed by Stryker Spine, Allendale, NJ 07401-1677. The MANTIS Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: degenerative disk disease, spodylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis and failed previous fusion.
Status
Terminated
Initiated
2009-04-30
Posted
2009-07-07
Terminated
2013-01-23
Reason
The Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod implantation and the split sphere may separate and fall into the patient.
Root cause
Device Design
Action
Stryker Spine issued an "Urgent Product Recall" notification dated April 29, 2009 via Federal Express to all Consignees. Users were instructed to quarantine all affected product and fax a copy of the Product Accountability Form to 201-760-8406. A Stryker sales representative will retrieve the quarantined product. For further information, contact Stryker Spine at 201-760-8206.
Quantity
387 in US
Distribution
Nationwide Distribution.
Lot, serial or model codes
Catalog Number: 48284051, all lots.   Mantis Rod inserter Shafts supplied separately, Catalog Number: 48284051.
510(k) numbers
["K061813"]
PMA numbers
not on file
Product code
MNH
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register