FDA Medical Device Recalls (openFDA)
Sybaritic, Inc: DERMALIFE Spa-Oceana 2G System. Part #E30008-01 - Spa-Oceana 2G; Assay. This is the Spa Oceana product along with a kit of products such as cleaner, aroma therapy concentrate, etc. Part #E30008 - Spa-Oceana 2G;220V,50/60HZ-15A. This is the Spa Oceana product along with a kit of products such as cleaner, aroma therapy concentrate, etc. that is slightly different than the kit of products included in Part #E30008-01 above.
- Recall number
- Z-1604-2009
- Recalling firm
- Sybaritic, Inc
- Product
- DERMALIFE Spa-Oceana 2G System. Part #E30008-01 - Spa-Oceana 2G; Assay. This is the Spa Oceana product along with a kit of products such as cleaner, aroma therapy concentrate, etc. Part #E30008 - Spa-Oceana 2G;220V,50/60HZ-15A. This is the Spa Oceana product along with a kit of products such as cleaner, aroma therapy concentrate, etc. that is slightly different than the kit of products included in Part #E30008-01 above.
- Status
- Terminated
- Initiated
- 2007-06-22
- Posted
- 2009-07-13
- Terminated
- 2012-07-05
- Reason
- A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Sybaritic, Inc. conducted a recall on DERMALIFE 2G equipment (SpaJet, Spa Oceana, Spa Feng Shui) because of IR Fan failure which resulted in the overheating and failure of the IR emitter and surrounding assembly.
- Root cause
- Device Design
- Action
- Sybaritic, Inc. issued a "Safety Alert" dated April 15, 2008 informing "All Owners/Users of DermaLife 2G Equipment" of the affected device. The letter advised consignees on device use until the mandatory Filed Correction Kit is delivered. For further questions, contact Sybaritic, Inc. at 1-800-445-8418.
- Quantity
- 86
- Distribution
- Worldwide Distribution -- US including states of CA, IA, MN, NE, NJ, NY, TX and WA and countries of Macedonia, Hungary, Greece, South Africa, Italy, Hong Kong, United Arab Emirates, Greece, Croatia, Australia, Kuwait, Turkey, Mexico, Belgium, England, Barbados, France, Japan, Lebanon, Moldova, Poland, Portugal, Romania, Russia, Spain and Thailand.
- Lot, serial or model codes
- Part E30008 SERIAL NUMBERS: G4061142236, G4060942128, G4061052135, G4061042136, G4070142436, G4070452585, G4070252464, G4061242317, G406124307, G406124311, G4061142281, G4070242474, G4070242486, G4070342498, G4070352528, G4061142257, G4061152258, G4061152273, G4070152376, G4070252458, G407044578, G4070342548, G4060842048, G4060852049, G4061152294, G4061252303, G4061252310, G4070242461, G4060952067, G40701424121, G4070152381, G4070152387, G4070342521, G4070352544, G4070552656, G4070552644, G406055898, G4060942068, G4061042145, G4061042214, G4061052149, G4061142244, G4061242326, G4061142299, G4070342506, G4070442577, G4060852053, G4060942074, G4061142259, G4061152228, G4061152240, G4061155248, G4070142363, G4070152400, G4070152408, G4070152437, G4070252463, G4070252473, G4070342517, G4070352495, G4070342529, G4060852019, G4061042186, G4061052212, G4061142224, G4061152220, G4070242448, G4070252442, G4070252484, G4070252487, G4070342491, G4070352505, G4070352509 and G4070352516; and Part E30008-01 SERIAL NUMBERS: G406052055, G4061252322, G4060852052, G4070142352, G4061052190, G4061042206, G4061242342, G4061042156, G4061052173, G4061052201, G4061252318, G4060852056 and G4060842058.
- 510(k) numbers
- K003916
- PMA numbers
- not on file
- Product code
- IMB
- Firm FEI
- 2131125
- Firm city
- Bloomington
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28