WillowWood Global Llc: Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha Lanyard - 700-AIS101, Alpha Interlock Suspension System with Alpha Lanyard w/Titanium Pyramid - 700-AIS107, Alpha Interlock Suspension System Lanyard Body ONLY

Recall number
Z-1594-2021
Recalling firm
WillowWood Global Llc
Product
Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha Lanyard - 700-AIS101, Alpha Interlock Suspension System with Alpha Lanyard w/Titanium Pyramid - 700-AIS107, Alpha Interlock Suspension System Lanyard Body ONLY
Status
Open, Classified
Initiated
2021-04-20
Posted
not on file
Terminated
not on file
Reason
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
Root cause
Process change control
Action
On 04/20/2021, Willow Wood issued an Urgent Medical Device Removal notice to customer via letter notifying them it was discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
Quantity
815 parts
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, KY, LA, MA, ME, MN, MO, MS, NJ, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and the countries of Germany, Netherlands
Lot, serial or model codes
Lot: 'xx' stamped on the outside of the Lanyard body manufactured form 04/01/2020 - 04/08/2021
510(k) numbers
not on file
PMA numbers
not on file
Product code
IPM
Firm FEI
1522166
Firm city
Mount Sterling
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register