Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Diagnostics Scarborough, Inc.: Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel nucleic acid assay system Component: No

Recall number
Z-1573-2025
Recalling firm
Abbott Diagnostics Scarborough, Inc.
Product
Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel nucleic acid assay system Component: No
Status
Open, Classified
Initiated
2025-03-04
Posted
2025-04-15
Terminated
not on file
Reason
The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.
Root cause
Process change control
Action
On March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers. Please complete the following actions, as applicable. 1. If you have impacted inventory in stock, then: " Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures. " Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot). " Please retain this letter for your records. 2. If you have forwarded the product listed above to others in your network, then: - Inform them of this Field Correction Notice, provide to them a copy of this notice and request they take the necessary action. 3. If you do not have impacted inventory in stock, then: - All product lots not identified in Appendix I can continue to be used. If you have questions regarding this information, please contact Sedgwick by phone at 888 671 8858 (available 8am 5pm EST, Monday - Friday). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178). If you have experienced any patient or user injury associated with this Field Correction Notice, please immediately report the event to your local area Customer Service. ****Update added 5/8/2025**** On April 29, 2025 FIELD CORRECTION NOTICE letters were sent to customers informing them of additional lots included in this recall.
Quantity
178056 eaches (147472 US, 31584 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom; Uruguay.
Lot, serial or model codes
Model 435-000; GTIN/DI 10811877010521; Lot 000M906790 Model 435-000; GTIN/DI 10811877010521; Lot 000M907271 Model 435-000; GTIN/DI 10811877010521; Lot 000M923531 Model 435-000; GTIN/DI 10811877010521; Lot 000M923879 Model 435-000; GTIN/DI 10811877010521; Lot 000M924414 Model 435-000; GTIN/DI 10811877010521; Lot 000M924430 Model 435-000; GTIN/DI 10811877010521; Lot 000M926642 Model 435-000; GTIN/DI 10811877010521; Lot 000M926679 Model 435-000; GTIN/DI 10811877010521; Lot 000M927161
510(k) numbers
K161375
PMA numbers
not on file
Product code
OCC
Firm FEI
1221359
Firm city
Scarborough
Firm state
ME
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register