Philips and Neusoft Medical Systems Co., Ltd.: NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Recall number
Z-1573-2016
Recalling firm
Philips and Neusoft Medical Systems Co., Ltd.
Product
NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Status
Terminated
Initiated
2016-03-24
Posted
2016-04-27
Terminated
2016-11-23
Reason
The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be
Root cause
Under Investigation by firm
Action
Affected consignees will be notified of the recall via letter on March 24, 2016. Mandatory Field Change Order and FCO kits will be released to all affected systems in June, 2016.
Quantity
37 units
Distribution
NC, OH, NE, SC, TX, LA, PR, MO, FL, CT
Lot, serial or model codes
Serial Numbers: N16E090002, N16E090003, N16E090007, N16E090008, N16E090010, N16E100023, N16E110023, N16E110026, N16E120016, N16E120023, N16E120052, N16E120053, N16E130012, N16E130038, N16E130045, N16E140008, N16E140007, N16E140012, N16E140017, N16E140020, N16E140029, N16E140056, N16E140057, N16E140058, N16E150004, N16E150014, N16E150013, N16E150015, N16E150025, N16E150024, N16E150027, N16E150038, N16E150039, N16E150040, N16E100017, N16E140030 and N16E140031
510(k) numbers
["K090173","K092742"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3003575350
Firm city
Sujiatun Shenyang
Firm state
not on file
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register