Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: 3/16 in. Curved Trocar with Wound Drain Curved Trocar with Wound Drain/Double Trocar with Wound Drain: The curved trocar wound drain has a curved metal trocar attached to one end of the flexible wound drain tube. There is also a polymeric trocar guard that provides protection from the sharp end of the trocar during handling. The double trocar wound drain has metal trocars on either side of the flexible wound drain tube. These are also equipped with trocar guards.

Recall number
Z-1570-2015
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
3/16 in. Curved Trocar with Wound Drain Curved Trocar with Wound Drain/Double Trocar with Wound Drain: The curved trocar wound drain has a curved metal trocar attached to one end of the flexible wound drain tube. There is also a polymeric trocar guard that provides protection from the sharp end of the trocar during handling. The double trocar wound drain has metal trocars on either side of the flexible wound drain tube. These are also equipped with trocar guards.
Status
Terminated
Initiated
2015-03-26
Posted
2015-05-01
Terminated
2015-12-31
Reason
Stryker Instruments is voluntarily recalling the CBCII Wound Drains and Ambulating Bulb for a potential sterility breach in the packaging
Root cause
Package design/selection
Action
An URGENT MEDICAL DEVICE NOTIFICATION letter was sent to all customers on 03/31/2015. The letter identified the affected product, the reason for the recall, the risk to health, and actions to be taken by the customer or user. All are asked to sign the Business Reply Form stating they have read the Notification and followed all instructions within. Customers were asked to immediately check all inventory and quarantine any affected product, return affected product to Stericycle for destruction using a return label provided by Stericycle. Upon receipt of affected product a credit will be issued to customer account.
Quantity
628 boxes (6,280 each)
Distribution
Worldwide Distribution. US nationwide, Australia, Great Britain, The Netherlands, Spain, Romania, Greece, Switzerland, South Africa, and Japan.
Lot, serial or model codes
Part Number: 0225-015-000 and lot numbers: 10083012 10349012 11224012 12151012 13092012 13196012 14034012 14246012 10125012 11014012 11257012 12286012 13119012 13248012 14127012 14263012 10162012 11059012 12039012 12341012 13135012 13274012 14165012 10273012 11108012 12075012 13010012 13172012 13326012 14177012 10309012 11160012 12080012 13065012 13191012 14021012 14231012
510(k) numbers
["K970714"]
PMA numbers
not on file
Product code
CAC
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register