Skip to the content

FDA Medical Device Recalls (openFDA)

Microgenics Corporation: Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and distributed by Microgenics Corporation, Fremont, CA MAS Bilirubin is intended for use as a consistent test sample of known concentration for monitoring assay conditions in many clinical laboratory determinations. Includes Bilirubin with patient serum specimens when assaying for any of the listed constituents. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument.

Recall number
Z-1563-2013
Recalling firm
Microgenics Corporation
Product
Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and distributed by Microgenics Corporation, Fremont, CA MAS Bilirubin is intended for use as a consistent test sample of known concentration for monitoring assay conditions in many clinical laboratory determinations. Includes Bilirubin with patient serum specimens when assaying for any of the listed constituents. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument.
Status
Terminated
Initiated
2013-05-13
Posted
2013-06-21
Terminated
2013-08-27
Reason
It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability.
Root cause
Under Investigation by firm
Action
ThermoFisher Scientific sent a Product Recall Notification letter dated May 13, 2013, by Certified US mail to all affected customer. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed 1. Determine if you are using or have inventory of the recalled product. 2. Discontinue use of the product immediately. 3. Evaluate quality controls. They may vary among laboratories and should be evaluated according to laboratory procedures. 4. Retain a copy of the letter for laboratory records. 5. If any of the affected lots of MAS Bilirubin had been forwarded to another laboratory, please provide a copy of the recall letter to them. 6. Complete and return the response form within 5 days. Letters also provide an alternative product - MAS Bilirubin part number BC103. Questions or concerns may be directed to Technical Services at 1-800-232-3342 or 510-979-5417.
Quantity
188
Distribution
Nationwide Distribution: US including the states of WI, UT, FL NY, PA, IL, CO, AZ and NJ.
Lot, serial or model codes
Catalog Number: 10014432;  Lot Numbers:  1MQ310, Expiration Date: 31 Dec 2013, 2CQ310, Expiration Date: 31 Mar 2014, 2KQ310, Expiration Date: 31 Oct 2014
510(k) numbers
["K031890"]
PMA numbers
not on file
Product code
JJX
Firm FEI
3010939897
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register