Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.

Recall number
Z-1551-2025
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.
Status
Open, Classified
Initiated
2025-03-11
Posted
2025-04-10
Terminated
not on file
Reason
Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.
Root cause
Software Design Change
Action
On March 11, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " For both Atellica CH and Atellica CI analyzers, perform the instructions in Appendix B to temporarily reduce the measuring interval. o Until the measuring interval is restored, track additional reagent consumption as a result of these actions to report to Siemens Healthineers for future reimbursement/credit. " Additionally, for Atellica CH Analyzers, perform the instructions in Appendix C to remove rules for flagging of No Calculation results and to install Atellica Solution Software version 1.29.0 or higher. " Complete and return the Field Correction Effectiveness Check form attached to this letter within 30 days. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution A follow-up communication will be provided when Customer Actions are no longer required. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Quantity
not on file
Distribution
Domestic: Nationwide Distribution. International: Australia, Austria, Bahamas, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark,¿Estonia, Finland, France, Germany, Greece,¿Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Netherlands, New Zealand, Norway, Oman,¿Pakistan, Poland, Portugal, Qatar, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, U.A.E., Uganda, United¿Kingdom, Uzbekistan, Vatikancity.¿ ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿¿¿¿ ¿ ¿¿¿¿ ¿¿¿¿¿¿ ¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿¿¿¿¿ ¿¿¿¿¿ ¿ ¿¿¿¿¿¿
Lot, serial or model codes
Siemens Material Number (SMN): 11537223; UDI-DI: 00630414610887; Lot numbers: All lots.
510(k) numbers
K221119
PMA numbers
not on file
Product code
DCN
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register