Skip to the content

FDA Medical Device Recalls (openFDA)

ZOLL Medical Corporation, World Wide Headquarters: ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll Medical Corporation, Chelmsford, MA. Intended for the purpose of converting ventricular fibrillation to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.

Recall number
Z-1547-2010
Recalling firm
ZOLL Medical Corporation, World Wide Headquarters
Product
ZOLL E Series Defibrillator/Pacemaker/Monitors, BLS Model, with software versions 4.XX, 6.XX and 7.0X. Zoll Medical Corporation, Chelmsford, MA. Intended for the purpose of converting ventricular fibrillation to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
Status
Terminated
Initiated
2010-03-24
Posted
2010-05-07
Terminated
2014-01-23
Reason
Device issued Shock Advised message but failed to auto-charge the defibrillator.
Root cause
Software design
Action
Zoll Medical Corporation issued an "Urgent Medical Device Correction" notification dated March 26, 2010 via certified mail to consignees who have received affected product. Customers will be contacted and given the option to have a software upgrade kit to update their devices or return devices to the firm to have the device upgraded. For further information, contact Zoll Medical Corporation 24/7 Technical Support at 1-800-348-9011 or 1-978-421-9460.
Quantity
1383 units
Distribution
Worldwide Distribution -- United States, Australia, Germany, Libyan Arab Jamahiriya, Malaysia, Singapore, United Arab Emirates and the United Kingdom.
Lot, serial or model codes
Software versions 4.XX, 6.XX and 7.0X in the E Series BLS Model.
510(k) numbers
K042007
PMA numbers
not on file
Product code
MKJ
Firm FEI
1220908
Firm city
Chelmsford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register