Beckman Coulter Inc: Cytomics FC 500 Flow Cytometry System with Data Innovations Instrument Manager, Part Numbers 626553 and 6605628 A system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.

Recall number
Z-1529-2008
Recalling firm
Beckman Coulter Inc
Product
Cytomics FC 500 Flow Cytometry System with Data Innovations Instrument Manager, Part Numbers 626553 and 6605628 A system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.
Status
Terminated
Initiated
2007-08-20
Posted
2008-08-08
Terminated
2012-10-15
Reason
Labeling provided by two integrated software systems are not clear enough to avoid potential demographic and sample type mismatches under certain conditions.
Root cause
Software design
Action
A Product Corrective Action (PCA) letter, dated August 16, 2007, informed the customers that the User documentation and Online Help for the FC500 with Data Innovations lnstrument Manager are not clear enough to avoid potential demographic and sample type mismatches under certain conditions. The letter asks the customers to share the information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. They are also to review all sample identification, demographics, data and results before reporting results out. A response form is included. Questions and concerns regarding the notification should be directed to Beckman Coulter Customer Service at 1-800-526-7694.
Quantity
141 units
Distribution
Worldwide Distribution --- including USA and country of Canada.
Lot, serial or model codes
All Software Versions
510(k) numbers
["K030828"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register