DePuy Orthopaedics, Inc.: DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.
- Recall number
- Z-1520-2013
- Recalling firm
- DePuy Orthopaedics, Inc.
- Product
- DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.
- Status
- Terminated
- Initiated
- 2013-05-03
- Posted
- 2013-06-07
- Terminated
- 2013-08-28
- Reason
- The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.
- Root cause
- Process control
- Action
- DePuy Synthes sent and Urgent Information - Medical Device Instrument Recall Notice letter dated May 3, 2013 via email to affected customers. The letter identified the affected product, problem, clinical implications, and potential adverse events. Arrangements will be made for all affected Glenosphere Orientation Guides (part #230795000/lot #5120443) to be returned. Customers were asked to return the response form via fax (574-371-4964). Product related questions should be directed to DePuy Synthes Joint Reconstruction (1-888-554-2482). Recall inquiries should be directed to Kim Earle, Recall Coordinator, 574-371-4917 (M-F; 8a.m.-5 p.m. EDT).
- Quantity
- 20 units
- Distribution
- Worldwide Distribution - USA Nationwide and countries of: Australia, Canada, Czech Republic, England, Germany, Switzerland and Turkey.
- Lot, serial or model codes
- Part Number 230795000/Lot Number 5120443
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWT
- Firm FEI
- 1818910
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28