Skip to the content

FDA Medical Device Recalls (openFDA)

Acumed LLC: The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between 35mm and 90mm. It is to be used in conjunction with commonly used orthopedic wire. 50.0 mm Tension Band Pin; Part Number 30-0097 70.0 mm Tension Band Pin; Part Number 30-0098 90.0 mm Tension Band Pin; Part Number 30-0099 50.0 mm Tension Band Kit; Part Number TB-1550K-S 70.0 mm Tension Band Kit; Part Number TB-1570K-S 90.0 mm Tension Band Kit; Part Number TB-1590K-S Product Usage: Acumed Tension Band Pin is intended to be used with orthopedic wire to address malleolar, patella, and olecranon fractures in tension band wiring procedures. This device is not intended for usage in the spine.

Recall number
Z-1516-2014
Recalling firm
Acumed LLC
Product
The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between 35mm and 90mm. It is to be used in conjunction with commonly used orthopedic wire. 50.0 mm Tension Band Pin; Part Number 30-0097 70.0 mm Tension Band Pin; Part Number 30-0098 90.0 mm Tension Band Pin; Part Number 30-0099 50.0 mm Tension Band Kit; Part Number TB-1550K-S 70.0 mm Tension Band Kit; Part Number TB-1570K-S 90.0 mm Tension Band Kit; Part Number TB-1590K-S Product Usage: Acumed Tension Band Pin is intended to be used with orthopedic wire to address malleolar, patella, and olecranon fractures in tension band wiring procedures. This device is not intended for usage in the spine.
Status
Terminated
Initiated
2014-03-26
Posted
2014-04-25
Terminated
2015-02-04
Reason
Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications.
Root cause
Component change control
Action
Acumed sent an Urgent Medical Device Recall letter dated March 26, 2014 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers are instructed to immediately stop using and/or distributing all lots identified in the letter. Return affected products as instructed in the letter. For questions contact Acumed Business Service via email at BusinessService@acumed.net or via phone at 877-627-9957.
Quantity
2544 unit in the US and 3584 units outside the US
Distribution
US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
Lot, serial or model codes
50.0 mm Tension Band Pin: Lot 243639 70.0 mm Tension Band Pin: Lots 241127, 242707, 243325, 243948. 90.0 mm Tension Band Pin: Lots 241128. 242708, 243327, 243949.  50.0 mm Tension Band Kit: Lots 216769, 217002, 217941, 219145, 219402, 222732, 223007, 224312, 229889, 232575, 235490, 237772, 244800, 246204, 247056.  70.0 mm Tension Band Kit: Lots 210290, 211349, 214574, 215405, 217004, 218273, 219230, 220978, 221503, 223008, 223477, 224066, 225554, 227193, 229394, 230287, 230952, 231971, 233742, 235502, 237773, 240512, 241185, 243617, 243945, 244801, 246777, 246779, 246938, 247108, 247109, 247493.  90.0 mm Tension Band Kit: Lots 215053, 217005, 218600, 218700, 221465, 222731, 223009, 223756, 225555, 228951, 231573, 236969, 237774, 243620, 244802, 246939, 246990, 247381.
510(k) numbers
K931244
PMA numbers
not on file
Product code
HRS
Firm FEI
1000125930
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register