Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.

Recall number
Z-1514-2014
Recalling firm
Elekta, Inc.
Product
HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.
Status
Terminated
Initiated
2014-04-17
Posted
2014-04-25
Terminated
2015-01-29
Reason
If screws are not properly tightened with a torque wrench they may become loose over time.
Root cause
Component design/selection
Action
Corrective Action #1 - Important Field Safety Notice 618 01 302 002 was issued April 2014. The Notice informs end users about the problem and the potential clinical impact. It also give a recommendation how the risk can be mitigated - inspect the screws and if loose contact Elekta Service to tighten the screws with 20 Nm. The notice also advises the customer that a modification kit will be released that introduces an improved screw solution. The solution will be implemented on all affected products when available. Corrective Action #2 Permanent Solution - Replace Current Screws - An improved screw has been designed and will be available for Elekta Field Service organization to replace on all affected products at the end of April. The new design solution provides a very long used thread depth, exceeding the standard recommended thread-depths by far. This design will prevent the screws from loosening.
Quantity
266
Distribution
Worldwide Distribution-USA including DC and the states of AZ, CA, CT, FL, GA, IL, HI, KY, MD, MA, MI, MS, MO, NJ, NY, NH, NC, ND, OH, OK, OR, PA, SC, TX, VA, WA, WI, and WY, and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Portugal, Russia, Saudi Arabia, South Korea, Span, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Lot, serial or model codes
Serial Numbers 42188, 42246, 42106, 42153, 42107, 42068, 42012, 42013, 42062, 42009, 42224, 42126, 3005, 42222, 3019, 3000, 42058, 42059, 42057, 42187, 42129, 42159, 3022, 3017, 42189, 42213, 42202, 42212, 42004, 4220, 42113, 42178, 42216, 42207, 2982, 2986, 2991, 42102, 42000, 42032, 42145, 42174, 42008, 42048, 42064, 42070, 42074, 42084, 42115, 42134, 42118, 42152, 42137, 42146, 3008, 2987, 3010, 3018, 42069, 42036, 3029, 42162, 42139, 42011, 42007, 42173, 42109, 42193, 3006, 42053, 42015, 42105, 42124, 42085, 42065, 42094, 42198, 2998, 3003, 3033, 3024, 2980, 42121, 2999, 42027, 42061, 42075, 42079, 42214, 3020, 42192, 42247, 42202, 42155, 42025, 42021, 42127, 42151, 42239, 42033, 42035, 42089, 42204, 42200, 2995, 3030, 42266, 42197, 42077, 42001, 42037, 42147, 42186, 42016, 42028, 42029, 42050, 42042, 42054, 42047, 42051, 42055, 42049, 42262, 42080, 42076, 42063, 42103, 42087, 42117, 42083, 42166, 42101, 42227, 42150, 42156, 42163, 42179, 42180, 42140, 42165, 42157,42234, 42169, 42182, 42183, 42236, 42195, 42223, 42256, 42276, 42260, 42231, 42250, 42243, 42240, 42272, 42235, 42249, 42254, 42245, 42265, 42268, 42267, 42253, 42264, 2993, 2997, 3004, 3031, 3002, 42226, 42258, 42114, 42081, 42090, 42093, 42133, 42181, 42039, 42116, 3016, 29677, 42010, 42073, 42176, 42164, 42031, 3028, 42041, 42038, 42022, 42034, 42088, 42104, 2976, 2979, 2988, 3011, 3015, 3021, 3027, 42238, 42018, 42135, 42252, 42185, 42273, 42060, 42024, 42006, 42082, 42194, 42263, 42232, 42199, 42170, 42161, 42078, 42158, 2985, 2994, 42120, 3032, 42056, 2969, 3034, 2975, 2972, 2984, 42191, 42225, 42228, 42023, 42143, 42122, 42208, 42230, 42211, 42026, 42206, 42066, 42052, 42014, 42218, 42030, 42040, 42244, 42043, 42044, 42045, 42046, 42123, 42255, 42141, 42125, 42072, 42112, 42210, 42017, 420196, 42020, 42067, 42071, 42097, 42100, 42131, 42144, 42136, 42148, 42130, 42111, 42067, 42171, 42203, 4210, 42205, 42096, 42154, 42241, 42229, 2981, 2989, 2971, 2990, 3007, 3013, 2992, 2973, 2978, 30098, 2974, 3014, 3012, 3025, 2996, 3026, 42274
510(k) numbers
["K072898"]
PMA numbers
not on file
Product code
JAI
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register