FDA Medical Device Recalls (openFDA)
Elekta, Inc.: HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.
- Recall number
- Z-1514-2014
- Recalling firm
- Elekta, Inc.
- Product
- HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.
- Status
- Terminated
- Initiated
- 2014-04-17
- Posted
- 2014-04-25
- Terminated
- 2015-01-29
- Reason
- If screws are not properly tightened with a torque wrench they may become loose over time.
- Root cause
- Component design/selection
- Action
- Corrective Action #1 - Important Field Safety Notice 618 01 302 002 was issued April 2014. The Notice informs end users about the problem and the potential clinical impact. It also give a recommendation how the risk can be mitigated - inspect the screws and if loose contact Elekta Service to tighten the screws with 20 Nm. The notice also advises the customer that a modification kit will be released that introduces an improved screw solution. The solution will be implemented on all affected products when available. Corrective Action #2 Permanent Solution - Replace Current Screws - An improved screw has been designed and will be available for Elekta Field Service organization to replace on all affected products at the end of April. The new design solution provides a very long used thread depth, exceeding the standard recommended thread-depths by far. This design will prevent the screws from loosening.
- Quantity
- 266
- Distribution
- Worldwide Distribution-USA including DC and the states of AZ, CA, CT, FL, GA, IL, HI, KY, MD, MA, MI, MS, MO, NJ, NY, NH, NC, ND, OH, OK, OR, PA, SC, TX, VA, WA, WI, and WY, and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Portugal, Russia, Saudi Arabia, South Korea, Span, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
- Lot, serial or model codes
- Serial Numbers 42188, 42246, 42106, 42153, 42107, 42068, 42012, 42013, 42062, 42009, 42224, 42126, 3005, 42222, 3019, 3000, 42058, 42059, 42057, 42187, 42129, 42159, 3022, 3017, 42189, 42213, 42202, 42212, 42004, 4220, 42113, 42178, 42216, 42207, 2982, 2986, 2991, 42102, 42000, 42032, 42145, 42174, 42008, 42048, 42064, 42070, 42074, 42084, 42115, 42134, 42118, 42152, 42137, 42146, 3008, 2987, 3010, 3018, 42069, 42036, 3029, 42162, 42139, 42011, 42007, 42173, 42109, 42193, 3006, 42053, 42015, 42105, 42124, 42085, 42065, 42094, 42198, 2998, 3003, 3033, 3024, 2980, 42121, 2999, 42027, 42061, 42075, 42079, 42214, 3020, 42192, 42247, 42202, 42155, 42025, 42021, 42127, 42151, 42239, 42033, 42035, 42089, 42204, 42200, 2995, 3030, 42266, 42197, 42077, 42001, 42037, 42147, 42186, 42016, 42028, 42029, 42050, 42042, 42054, 42047, 42051, 42055, 42049, 42262, 42080, 42076, 42063, 42103, 42087, 42117, 42083, 42166, 42101, 42227, 42150, 42156, 42163, 42179, 42180, 42140, 42165, 42157,42234, 42169, 42182, 42183, 42236, 42195, 42223, 42256, 42276, 42260, 42231, 42250, 42243, 42240, 42272, 42235, 42249, 42254, 42245, 42265, 42268, 42267, 42253, 42264, 2993, 2997, 3004, 3031, 3002, 42226, 42258, 42114, 42081, 42090, 42093, 42133, 42181, 42039, 42116, 3016, 29677, 42010, 42073, 42176, 42164, 42031, 3028, 42041, 42038, 42022, 42034, 42088, 42104, 2976, 2979, 2988, 3011, 3015, 3021, 3027, 42238, 42018, 42135, 42252, 42185, 42273, 42060, 42024, 42006, 42082, 42194, 42263, 42232, 42199, 42170, 42161, 42078, 42158, 2985, 2994, 42120, 3032, 42056, 2969, 3034, 2975, 2972, 2984, 42191, 42225, 42228, 42023, 42143, 42122, 42208, 42230, 42211, 42026, 42206, 42066, 42052, 42014, 42218, 42030, 42040, 42244, 42043, 42044, 42045, 42046, 42123, 42255, 42141, 42125, 42072, 42112, 42210, 42017, 420196, 42020, 42067, 42071, 42097, 42100, 42131, 42144, 42136, 42148, 42130, 42111, 42067, 42171, 42203, 4210, 42205, 42096, 42154, 42241, 42229, 2981, 2989, 2971, 2990, 3007, 3013, 2992, 2973, 2978, 30098, 2974, 3014, 3012, 3025, 2996, 3026, 42274
- 510(k) numbers
- ["K072898"]
- PMA numbers
- not on file
- Product code
- JAI
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28