FDA Medical Device Recalls (openFDA)
Datascope Corp.: Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65
- Recall number
- Z-1506-2021
- Recalling firm
- Datascope Corp.
- Product
- Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65
- Status
- Open, Classified
- Initiated
- 2021-03-22
- Posted
- 2021-04-28
- Terminated
- not on file
- Reason
- There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
- Root cause
- Software design
- Action
- URGENT MEDICAL DEVICE CORRECTION notification letters dated 3/22/21 were distributed to customers. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL CARDIOSAVE HYBRID and CARDIOSAVE RESCUE IABP USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Corrective Action: Datascope/Getinge is currently developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. Please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to cardiosaveripple20.act@getinge.com or by faxing the form to 1-800-506-1644.
- Quantity
- 10087 total
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Lot, serial or model codes
- All Systems. UDI: 10607567111117 10607567109008 10607567111940 10607567109107 10607567108421 10607567108438 10607567108391 10607567108414 10607567113432
- 510(k) numbers
- ["K112372","K151254","K163542"]
- PMA numbers
- not on file
- Product code
- DSP
- Firm FEI
- 3001418283
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28