Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corp.: Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65

Recall number
Z-1506-2021
Recalling firm
Datascope Corp.
Product
Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65
Status
Open, Classified
Initiated
2021-03-22
Posted
2021-04-28
Terminated
not on file
Reason
There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
Root cause
Software design
Action
URGENT MEDICAL DEVICE CORRECTION notification letters dated 3/22/21 were distributed to customers. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL CARDIOSAVE HYBRID and CARDIOSAVE RESCUE IABP USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Corrective Action: Datascope/Getinge is currently developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. Please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to cardiosaveripple20.act@getinge.com or by faxing the form to 1-800-506-1644.
Quantity
10087 total
Distribution
Worldwide distribution - US Nationwide distribution.
Lot, serial or model codes
All Systems. UDI: 10607567111117 10607567109008 10607567111940 10607567109107 10607567108421 10607567108438 10607567108391 10607567108414 10607567113432
510(k) numbers
["K112372","K151254","K163542"]
PMA numbers
not on file
Product code
DSP
Firm FEI
3001418283
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register