Skip to the content

FDA Medical Device Recalls (openFDA)

Turbett Surgical, Inc.: Turbett Surgical Container, TS1500

Recall number
Z-1501-2022
Recalling firm
Turbett Surgical, Inc.
Product
Turbett Surgical Container, TS1500
Status
Terminated
Initiated
2022-04-14
Posted
2022-08-04
Terminated
2025-02-20
Reason
Devices were subject to unapproved rework processes.
Root cause
Process control
Action
The firm distributed recall notification to hospitals via email on 04/14/2022. The recall notification identifies the reason for recall of specific TS1500 units to be that the repair process employed could restrict the customer's ability to clean and visually check the device for cleanliness. This can be identified by the customer by evidence of bioburden or particulate buildup in the seams of the device. Inadequate or incomplete cleaning of the unit may compromise the sterility of the load. The firm asks customers to immediately discontinue use of the device and complete the provided Return Response Form to Quality@Turbettsurgical.com. Turbett will contact customers to coordinate the replacement of the unit and correction of the recalled device. Customers with any questions can call 585-755-0133x106 from Monday through Friday, from 8:00 AM to 4:30 PM.
Quantity
7 units
Distribution
US Nationwide Nationwide: CO, FL, NC and PA.
Lot, serial or model codes
UDI-DI (GTIN): 00859464007113; Lot No./Serial No.: 100036/2017061015, 100036/2017061016, 100036/2017061017, 100036/2017061019, 100036/2017061023, 100036/2017061025, 100065/2017121001 and 100081/2018031004.
510(k) numbers
["K153025"]
PMA numbers
not on file
Product code
FRG
Firm FEI
3012421755
Firm city
Victor
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register