Skip to the content

FDA Medical Device Recalls (openFDA)

PerkinElmer LAS Inc: Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000.

Recall number
Z-1499-04
Recalling firm
PerkinElmer LAS Inc
Product
Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000.
Status
Terminated
Initiated
2004-09-08
Posted
2004-09-30
Terminated
2008-10-29
Reason
There is an increase in the number of normal specimen results below the lowest calibrator and based on a report that the normal controls were running low, sometimes out of the specified range.
Root cause
Other
Action
The firm contacted their consignees by letter sent 9/8/04.
Quantity
49 kits
Distribution
The kits were distributed to 5 consignees, located in Indianapolis, IN; Jefferson City, MO; Austin, TX; Madison, WI; and Minneapolis, MN.
Lot, serial or model codes
Lot #244468
510(k) numbers
not on file
PMA numbers
not on file
Product code
JIA
Firm FEI
1522510
Firm city
Norton
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register