FDA Medical Device Recalls (openFDA)
Diagnostic Hybrids, Inc.: MicroVue CIC-C1q EIA, Model A001. An enzyme immunoassay for the detection of circulating immune complexes in human serum or plasma. The assay uses highly pure, functional human C1q coated in the solid phase to capture immune complexes (CIC). In the first stage, CIC in the diluted patient samples and HAGG in the controls and standards are dispensed into the C1q coated assay wells. After incubation, unbound material is removed in a washing step and a ready to use conjugate (goat anti-human Ig-HRP) is added. After a second incubation, unbound conjugate is washed away. After addition of a substrate and a short incubation interval, the quantity of CIC in the sample (g Eq/ml) can be determined by comparison to a standard curve.
- Recall number
- Z-1496-2018
- Recalling firm
- Diagnostic Hybrids, Inc.
- Product
- MicroVue CIC-C1q EIA, Model A001. An enzyme immunoassay for the detection of circulating immune complexes in human serum or plasma. The assay uses highly pure, functional human C1q coated in the solid phase to capture immune complexes (CIC). In the first stage, CIC in the diluted patient samples and HAGG in the controls and standards are dispensed into the C1q coated assay wells. After incubation, unbound material is removed in a washing step and a ready to use conjugate (goat anti-human Ig-HRP) is added. After a second incubation, unbound conjugate is washed away. After addition of a substrate and a short incubation interval, the quantity of CIC in the sample (g Eq/ml) can be determined by comparison to a standard curve.
- Status
- Terminated
- Initiated
- 2017-04-04
- Posted
- not on file
- Terminated
- 2020-08-28
- Reason
- There is a potential for the slope of the standard curve to fail to meet the assay validation requirements for the assay.
- Root cause
- Other
- Action
- Customers were notified via letter on 4/4/17. Customers were asked to review their current inventory for the specified lot and complete and return the response form. If the product was further distributed, customers were instructed to provide them with the notification letter.
- Quantity
- 101 kits
- Distribution
- The products were distributed to the following US states: NY, TX, CO, UT. The products were distributed to the following foreign countries: Switzerland.
- Lot, serial or model codes
- Lot 084007
- 510(k) numbers
- K862649
- PMA numbers
- not on file
- Product code
- DAK
- Firm FEI
- 1000122536
- Firm city
- Athens
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as