Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Spectrum IQ - Infusion Pump - Product Usage: intended to be used for the controlled administration of fluids.

Recall number
Z-1487-2021
Recalling firm
Baxter Healthcare Corporation
Product
Spectrum IQ - Infusion Pump - Product Usage: intended to be used for the controlled administration of fluids.
Status
Terminated
Initiated
2021-03-19
Posted
not on file
Terminated
2022-07-15
Reason
Potentially defective component in the AC power adapter which may fail over time.
Root cause
Device Design
Action
On 3/19/2020 the firm sent a letter to its customers with the following instructions: 1. Operators may continue to use the Spectrum IQ Infusion Pumps until the correction is completed. If using an impacted pump to infuse critical medication such as vaso-active medications, ensure there is a back-up pump available in case the power adapter fails. 2. A local Baxter service representative will contact your facility to provide a list of affected serial numbers at your facility and to coordinate power adapter replacement. Your facility will be receiving this power adapter from Baxter at no charge. 3. Upon receipt of the new power adapters, refer to the Spectrum IQ Installation and Maintenance Manual (41019v900 Rev F, Section 11.2) for instructions on how to remove and replace the power adapter on the impacted pumps. The manual can be found on Baxters Technical Service Portal at https://service.baxter.com. 4. After the installation of the new power adapters, please return the original power adapters to Baxter using the packaging from the replacement power adapter and the provided return label. 5. If you received this notification directly from Baxter, complete the enclosed Baxter Customer Reply Form and return it to Baxter by e-mailing it to fca@baxter.com. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 6. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 7. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 8. If you are a de
Quantity
408 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CT, FL, IA, IN, LA, MA, MD, ME, MN, MO, MS, MT, NC, NJ, NY, OH, PA, RI, SC, TN, TX and the country of Puerto Rico.
Lot, serial or model codes
Product Code: 3570009 Serial Numbers: 3008222, 3539596, 3546215, 3586366, 3619652, 3622541, 3623709, 3624381, 3624740, 3625247, 3011477, 3539777, 3547110, 3586582, 3619759, 3622688, 3623721, 3624383, 3624744, 3625266, 3012363, 3540327, 3547708, 3586798, 3619767, 3622691, 3623728, 3624393, 3624749, 3625275, 3013708, 3540486, 3548756, 3587371, 3620063, 3622721, 3623737, 3624396, 3624761, 3625300, 3014132, 3541062, 3549098, 3587774, 3620333, 3622773, 3623754, 3624409, 3624763, 3625308, 3014459, 3541163, 3552089, 3588483, 3620532, 3622779, 3623789, 3624414, 3624777, 3625314, 3500621, 3541326, 3553030, 3589274, 3620606, 3622791, 3623801, 3624415, 3624784, 3625370, 3500772, 3541365, 3554960, 3590805, 3620620, 3622849, 3623843, 3624446, 3624805, 3625422, 3502812, 3541546, 3557201, 3597269, 3620626, 3622888, 3623873, 3624449, 3624818, 3625437, 3522442, 3541581, 3560533, 3598374, 3620699, 3622889, 3623877, 3624473, 3624824, 3625443, 3525471, 3541732, 3560780, 3599028, 3620766, 3623048, 3623878, 3624493, 3624825, 3625512, 3526276, 3541857, 3562368, 3600400, 3620769, 3623157, 3623886, 3624497, 3624826, 3625548, 3526801, 3541923, 3563542, 3601482, 3620775, 3623241, 3623895, 3624498, 3624827, 3625634, 3526869, 3541932, 3564532, 3601498, 3620790, 3623244, 3623921, 3624539, 3624828, 3625648, 3527443, 3541971, 3566208, 3601772, 3620887, 3623295, 3623962, 3624541, 3624830, 3625684, 3527832, 3542101, 3566535, 3602192, 3620905, 3623349, 3623985, 3624553, 3624835, 3625685, 3528039, 3542226, 3566790, 3602246, 3621090, 3623416, 3623991, 3624567, 3624838, 3625694, 3528084, 3542278, 3566871, 3607943, 3621234, 3623420, 3624034, 3624572, 3624839, 3625695, 3528271, 3542683, 3569103, 3608133, 3621315, 3623434, 3624069, 3624573, 3624853, 3625698, 3528426, 3542739, 3571487, 3609850, 3621351, 3623441, 3624155, 3624590, 3624905, 3625801, 3528462, 3543043, 3572131, 3609928, 3621354, 3623444, 3624162, 3624591, 3624924, 3625804, 3528535, 3543442, 3572541, 3610434, 3621459, 3623447, 3624173, 3624597, 3624925, 3625805, 3528601, 3543477, 3572696, 3610714, 3621534, 3623452, 3624185, 3624622, 3624930, 3625831, 3530776, 3543509, 3572824, 3611054, 3621591, 3623455, 3624214, 3624623, 3624934, 3625836, 3530964, 3543792, 3572909, 3611702, 3621611, 3623460, 3624221, 3624624, 3624937, 3625913, 3531978, 3543909, 3573095, 3612451, 3621684, 3623493, 3624230, 3624629, 3624966, 3625915, 3532405, 3544169, 3573194, 3614169, 3621705, 3623524, 3624243, 3624639, 3624968, 3625918, 3533275, 3544200, 3573272, 3614468, 3621786, 3623576, 3624252, 3624658, 3624971, 3625921, 3536232, 3544466, 3573666, 3616985, 3621848, 3623584, 3624259, 3624660, 3624982, 3625981, 3537657, 3544530, 3573892, 3617094, 3622007, 3623595, 3624280, 3624661, 3625003, 3626030, 3538453, 3544540, 3573916, 3617407, 3622011, 3623616, 3624283, 3624670, 3625004, 3626032, 3538557, 3544657, 3574481, 3617538, 3622077, 3623636, 3624288, 3624678, 3625025, 3626116, 3538693, 3544745, 3575642, 3617745, 3622086, 3623647, 3624290, 3624700, 3625044, 3626198, 3538779, 3544805, 3575858, 3617883, 3622133, 3623649, 3624297, 3624701, 3625115, 3626232, 3538840, 3544914, 3577824, 3618471, 3622146, 3623653, 3624303, 3624719, 3625132, 3626252,  3538847, 3545023, 3577991, 3618800, 3622148, 3623657, 3624306, 3624721, 3625156, 3626257, 3538867, 3545037, 3578563, 3618934, 3622159, 3623659, 3624339, 3624724, 3625158, 3626515, 3539013, 3545101, 3578574, 3618954, 3622201, 3623665, 3624349, 3624727, 3625186, 3628102, 3539331, 3545730, 3578811, 3619059, 3622364, 3623668, 3624358, 3624729, 3625188, 3629743, 3539375, 3545967, 3579301, 3619592, 3622429, 3623669, 3624379, 3624735, 3625234, 3539423, 3546068, 3585654, 3619625, 3622450, 3623691, 3624380, 3624736, 3625240.
510(k) numbers
["K173084"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1314492
Firm city
Medina
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register