Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien Llc: TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808

Recall number
Z-1467-2022
Recalling firm
Covidien Llc
Product
TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808
Status
Open, Classified
Initiated
2022-06-07
Posted
not on file
Terminated
not on file
Reason
Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.
Root cause
Device Design
Action
Medtronic issued Urgent Medical Device Recall letter on 6/7/22 (Field Action #: FA1259) via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take: "Identify and quarantine all affected product. The enclosed Customer Notification DetailReport includes affected product for your facility. " Return all affected product in your inventory to Medtronic. a. Product purchased directly from Medtronic: rs.covidienfeedbackcustomerservice@medtronic.com b. Product purchased through distributor: rs.gmbfcamitg@medtronic.com " Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Complete the enclosed Customer Confirmation Form and email it to rs.covidienfeedbackcustomerservice@medtronic.com " If you have questions or concerns regarding this recall, please contact Medtronic CustomerService at 800-854-3570.
Quantity
33 units
Distribution
AZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom Ireland Netherlands Reunion Spain
Lot, serial or model codes
UDI-Device Identifier (GTIN/UPN) 10884521749931 Serial Number: YF1120, YF1126, YF1127, YF1128, YF1129, YF1130, YF1131, YF1133, YF1134, YF1135, YF1137, YF1138, YF1139, YF1140, YF1141, YF1142, YF1150, YF1152, YF1154, YF1156, YF1157, YF1158, YF1161, YF1163, YF1166, YF1167, YF1168, YF1169, YF1172, YF1175, YF1176, YF1177, YF1182
510(k) numbers
["K031787","K132015"]
PMA numbers
not on file
Product code
HIH
Firm FEI
1282497
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register