Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien LP: VersaOne Reusable Positioning Cannula 12mm Standard, Model Number RC12STS

Recall number
Z-1465-2023
Recalling firm
Covidien LP
Product
VersaOne Reusable Positioning Cannula 12mm Standard, Model Number RC12STS
Status
Open, Classified
Initiated
2023-03-22
Posted
2023-04-26
Terminated
not on file
Reason
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
Root cause
Process control
Action
An URGENT MEDICAL DEVICE RECALL notification letter dated March 2023 was sent to customers. Customer Actions: Our records show that your facility has received the impacted product. Medtronic requests that you take the following actions: " Identify and quarantine any used and unused impacted product. Please refer to Attachment A: Impacted Serial Number Lists for a list of affected products. For verifying the respective product and serial number on the device and packaging please refer to Attachment B. " Sterilize all used devices as per your standard decontamination protocol prior to returning used cannulas. Utilization of appropriate personal protective equipment is advisable during decontamination of the device. " Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form. " Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information. " Forward this notification to any entities within your organization or to any other organization where the potentially affected product has been transferred or further distributed. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: " Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) " Medtronic Technical Services at 1-800-962-9888, option 2 to provide information regarding those events. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Field Representative.
Quantity
1 units
Distribution
Worldwide - US Nationwide distribution in the states of NC and the countries of Belgium, Denmark, France, Germany, India, Italy, Netherlands, Spain, Sweden.
Lot, serial or model codes
UDI-DI: 10884521784062; Serial Numbers: C21RXG0005
510(k) numbers
["K201805"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
1282497
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register