Optovue, Inc.: Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended for in vivo imaging.

Recall number
Z-1441-2013
Recalling firm
Optovue, Inc.
Product
Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended for in vivo imaging.
Status
Terminated
Initiated
2013-05-14
Posted
2013-05-31
Terminated
2013-10-24
Reason
Colors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international) are different from those in the individual iWellness ID and OS reports.
Root cause
Software design
Action
Optovue sent a Field Correction Notice letter dated May 16, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Quantity
~329 users
Distribution
Distributed Nationwide and in Canada.
Lot, serial or model codes
iVue software version 3.0 and 3.1 3.1 is international version, only distributed to Canada but not currently installed in any Canadian facilities.
510(k) numbers
K121739
PMA numbers
not on file
Product code
HLI
Firm FEI
3005950902
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register