Philips Healthcare Inc.: CareVue Chart Release C.O Part Number: 862246 Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
- Recall number
- Z-1436-2009
- Recalling firm
- Philips Healthcare Inc.
- Product
- CareVue Chart Release C.O Part Number: 862246 Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
- Status
- Terminated
- Initiated
- 2009-05-05
- Posted
- 2009-05-27
- Terminated
- 2012-03-06
- Reason
- Software: A software defect under certain circumstances could lead to the unintended administration of a duplicate dose of a scheduled medication within a single 24 hour period
- Root cause
- Software design
- Action
- Philips Medical notified accounts by letter on 5/6/09 the Urgent - Medical Device Correction notification. While customers await the correction of their devices, customers are given the following instructions: Clinicians should not rely solely on the worklist within CareVue Chart to determine actionable orders for the patient until their software is upgraded. Clinical users should always use the medication administration record (MAR) and the administration record in determining the care their patients receive in addition to the critical care worklist. Philips is currently developing a correction that will be available at no additional charge. A Philips representative will notify customers when the software correction becomes available. If further assistance is needed, customers are to contact the Philips Customer Care Center at 1-800-722-9377.
- Quantity
- 103 units
- Distribution
- Worldwide Distribution -- USA including states of CA, FL, MA, OH, OK, RI, VT, and WV and countries of Canada, Australia, Austria, Belgium, Denmark, Finland, France, Germany, Italy, Malaysia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, and the United Kingdom.
- Lot, serial or model codes
- Serial Numbers: 4621A00101, 4621A00102, 4621A00103, 4621A00104, 4621A00105, 4621A00106, 4621A00107, 4621A00108, 4621A00109, 4621A00110, 4621A00111, 4621A00112, 4621A00113, 4621A00114, 4621A00115, 4621A00116, 4621A00117, 4621A00118, 4621A00119, 4621A00120, 4621A00121, 4621A00122, 4621A00123, 4621A00124, 4621A00125, 4621A00126, 4621A00127, 4621A00128, 4621A00129, 4621A00130, 4621A00131, 4621A00132, 4621A00133, 4621A00134, 4621A00135, 4621A00136, 4621A00137, 4621A00138, 4621A00139, 4621A00140, 4621A00141, 4621A00142, 4621A00143, 4621A00144, 4621A00145, 4621A00146, 4621A00147, 4621A00148, 4621A00149, 4621A00150, 4621A00151, 4621A00152, 4621A00153, 4621A00154, 4621A00155, 4621A00156, 4621A00157, 4621A00158, 4621A00159, 4621A00160, 4621A00161, 4621A00162, 4621A00163, 4621A00164, 4621A00165, 4621A00166, 4621A00167, 4621A00168, 4621A00169, 4621A00170, 4621A00171, 4621A00172, 4621A00193, 4621A00194, 4621A00195, 4621A00196, 4621A00197, 4621A00198, 4621A00199, 4621A00200, 4621A00201, 4621A00202, 4621A00203, 4621A00204, 4621A00205, 4621A00206, 4621A00207, 4621A00208, 4621A00209, 4621A00210, 4621A00211, 4621A00213, 4621A00214, 4621A00217, 4621A00218, 4621A00219, 4621A00220, 4621A00221, 4621A00222, 4621A00223, and 4621A00224.
- 510(k) numbers
- ["K992636"]
- PMA numbers
- not on file
- Product code
- DXJ
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28