Cordis Corporation: Cordis, AVANTI+ Introducer, STERILE EO, For one use only. Catalog 402656X Lot # 15076369. Catalog 504606X Lot # 15075717. Catalog 504656X Lot # 15076362, 15076363, 15076364.
- Recall number
- Z-1421-2010
- Recalling firm
- Cordis Corporation
- Product
- Cordis, AVANTI+ Introducer, STERILE EO, For one use only. Catalog 402656X Lot # 15076369. Catalog 504606X Lot # 15075717. Catalog 504656X Lot # 15076362, 15076363, 15076364.
- Status
- Terminated
- Initiated
- 2010-01-28
- Posted
- 2010-04-20
- Terminated
- 2012-07-27
- Reason
- Sterility compromised: Cordis found a pin hole in one of the packaging tray cavities.On January 28, 2010 Cordis Corporation initiated a recall on their AVANTI + Introducer. During routine inspection, a visual irregularity was detected in packing trays.
- Root cause
- Package design/selection
- Action
- Cordis initiated the recall January 28, 2010. (The packets were sent overnight on January 28, 2010, with first receipt by customers January 29, 2010.) Immediate corrective action was taken through an inspection and removal of all affected products.
- Quantity
- 1850 units
- Distribution
- Nationwide.
- Lot, serial or model codes
- Catalog 402656X Lot # 15076369. Catalog 504606X Lot # 15075717. Catalog 504656X Lot # 15076362, 15076363, and 15076364.
- 510(k) numbers
- ["K945616"]
- PMA numbers
- not on file
- Product code
- DYB
- Firm FEI
- 1016427
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28