Microvention, Inc.: MicroVention Terumo Microplex, Platinum Coil System, Endovascular Embolization Coil, VFC, Sterile, Rx, REF numbers MV-00306VFC, MV-00315VFC, MV-00310VFC, MV-00630VFC, VFC010303-V, VFC030606-V, VFC030610-V, VFC030615-V, VFC061020-V, VFC061030-V, VFC101530-V, VFC101540-V, VFC152040-V, and VFC152060-V. Product Usage: . The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
- Recall number
- Z-1406-2020
- Recalling firm
- Microvention, Inc.
- Product
- MicroVention Terumo Microplex, Platinum Coil System, Endovascular Embolization Coil, VFC, Sterile, Rx, REF numbers MV-00306VFC, MV-00315VFC, MV-00310VFC, MV-00630VFC, VFC010303-V, VFC030606-V, VFC030610-V, VFC030615-V, VFC061020-V, VFC061030-V, VFC101530-V, VFC101540-V, VFC152040-V, and VFC152060-V. Product Usage: . The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
- Status
- Terminated
- Initiated
- 2019-11-22
- Posted
- not on file
- Terminated
- 2022-01-26
- Reason
- The devices may be missing the implant coil.
- Root cause
- Process design
- Action
- The recalling firm issued letters beginning 11/22/2019 dated 11/26/2019 via email using different flags according to the region the foreign consignee was located, i.e. 'URGENT: FIELD SAFETY CORRECTION ACTION" (for EMEA), "URGENT: MEDICAL DEVICE RECALL" (for Japan), or "FIELD SAFETY ADVISORY NOTICE" (for Rest of World).
- Quantity
- 666 units
- Distribution
- All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
- Lot, serial or model codes
- Catalog/lot numbers: MV-00306VFC - 1904105NZ and 1905085NZ; MV-00310VFC - 1903045TZ, 1904035TZ, 1904085NZ, AND 1905065NZ; MV-00315VFC - 1903045TZ, 1903155NZ, and 1904055YV; MV-00630VFC - 1904085FV and 1905085FV; VFC010303-V - 1903065WV, 1904015WV, 1904105WV, 1904155WV, 1904175WV, 1904225WV, 1904245WV, 1904265WV, 1905065WV, 1905105WV, 1905155WV, and 1905175WV; VFC030606-V - 1903045WV, 1903065WV, 1903085PV, 1903205WV, 1904015WV, 1904035WV, 1904055WV, 1904105WV, 1904155PV, 1904175UV, 1904175WV, 1904225WV, 1904245UV, 1904245WV, 1904265WV, 1905065WV, and 1905175WV; VFC030610-V - 1903015WV, 1903065WV, 1904015WV, 1904035WV, 1904055WV, 1904155WV, 1904175UV, 1904175WV, 1904225UV, 1904225WV, 1904245UV, 1904245WV, 1904265WV, 1905065WV, 1905085WV, 1905105WV, 1905135WV, 1905155PV, 1905175WV, and 1905225WV; VFC030615-V - 1903015WV, 1903045WV, 1903065PV, 1903065WV, 1903225UV, 1903225ZV, 1904015WV, 1904055WV, 1904085WV, 1904245WV, 1904265WV, 1905025WV, 1905035WV, 1905085PV, 1905085WV, and 1905175WV; VFC061020-V - 1903045WV, 1903065WV, 1903205WV, 1904225WV, 1904265WV, 1905025WV, 1905065WV, 1905085WV 1905175WV, and 1905225WV; VFC061030-V - 1903065WV, 1904015PV, 1904015WV, 1904055WV, 1904225WV, 1904245WV, 1904265WV, 1905035WV, 1905065PV, and 1903115RV; VFC101530-V - 1903135WV, 1904265WV, 1905025WV, 1905105WV, 1905175WV, and 1905205WV; VFC101540-V - 1903045WV, 1903185WV, 1904265WV, 1905025WV, 1905035WV, and 1905225WV; VFC152040-V - 1903135RV; and VFC152060-V - 1903085RV, 1904265RV, and 1905135RV.
- 510(k) numbers
- ["K050954","K131948","K132317","K132952","K153594"]
- PMA numbers
- not on file
- Product code
- HCG
- Firm FEI
- 3003523199
- Firm city
- Aliso Viejo
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28