Elekta, Inc.: Elekta Synergy XVI Medical Linear Accelerator. Model Numbers R3.5, R4.0 and R4.2. Elekta, Inc. Norcross, GA 30092. Intended for use as a radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-1405-2010
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Synergy XVI Medical Linear Accelerator. Model Numbers R3.5, R4.0 and R4.2. Elekta, Inc. Norcross, GA 30092. Intended for use as a radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Status
- Terminated
- Initiated
- 2008-12-12
- Posted
- 2010-04-16
- Terminated
- 2010-12-28
- Reason
- It is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing.
- Root cause
- Software design
- Action
- Elekta, Inc. issued an "Important Notice" dated December 12, 2008. Customers were informed of the affected device and instructed on proper steps until an update is provided. For further information, contact Elekta, Inc. at 1-770-300-9725.
- Quantity
- 76 units
- Distribution
- Nationwide Distribution -- AL, AZ, CO, CT, FL, GA, ID, KS, KY, MA, MI, MN, MS, NE, NY, OH, OK, OR, PA, TN, TX, VA, WA and WV.
- Lot, serial or model codes
- Serial Numbers: 151008, 151024, 151025, 151033, 151038, 151051, 151053, 151055, 151059, 151073, 151074, 151076, 151080, 151084, 151093, 151104, 151110, 151130, 151134, 151141, 151143, 151151, 151154, 151155, 151156, 151157, 151160, 151167, 151168, 151173, 151176, 151178, 151216, 151228, 151229, 151238, 151245, 151250, 151253, 151256, 151259, 151260, 151266, 151267, 151269, 151298, 151301, 151305, 151309, 151317, 151325, 151326, 151327, 151328, 151340, 151352, 151359, 151360, 151362, 151363, 151369, 151371, 151387, 151394, 151398, 151411, 151416, 151421, 151422, 151423, 151430, 151433, 151435, 151441, 151446 and 151479.
- 510(k) numbers
- K051932
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28