R & D Systems, Inc.: Quantikine IVD Erythropoietin Human Serum Controls for use as quantitative controls for the determination of Erythropoietin concentrations in human serum and plasma. IFU, Part Number 750095

Recall number
Z-1390-2022
Recalling firm
R & D Systems, Inc.
Product
Quantikine IVD Erythropoietin Human Serum Controls for use as quantitative controls for the determination of Erythropoietin concentrations in human serum and plasma. IFU, Part Number 750095
Status
Completed
Initiated
2022-06-14
Posted
2022-07-15
Terminated
not on file
Reason
The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards Institution number 2797 (BSI-2797) for product inserts 750095.28 and 750095.29. The erroneous label CE2797 were in the IFU insert revision 750095.28 for the 408 affected lots that were shipped. No product was packed with erroneous insert revision 750095.29. Since the product is currently CE marked as a Class I - self-certified product under the EU In Vitro Diagnostics Directive, the Notified Body designation should not be specified. As part of the corrective action, the firm has notified the customers who received the products with erroneous labeling, and is conducting the replacements with the correct product insert containing the Insert Revision 750095.30
Root cause
Process control
Action
The firm notified its consignees on 06/14/2022 by email. Distributors were directed to notify their customers. Users were directed to destroy any inventory.
Quantity
400
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, MN, UT, OH, MA, TX, PR, PA, NJ, CA, WI, LA, Puerto Rico, and the countries of CANADA, UNITED KINGDOM, ISRAEL, MEXICO, THAILAND, CHINA, HONG KONG, AUSTRALIA, INDONESIA.
Lot, serial or model codes
Lot Code: UDI 00815762021671, Lot Codes: P301801; P304262; P306132; P310038; P314685; P320216; P322531 Expiration Date: 08/04/2023
510(k) numbers
["K931756"]
PMA numbers
not on file
Product code
GGT
Firm FEI
2182501
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register