Angiodynamics, Inc.: NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN: H787204001110

Recall number
Z-1381-2023
Recalling firm
Angiodynamics, Inc.
Product
NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN: H787204001110
Status
Open, Classified
Initiated
2023-03-02
Posted
2023-04-12
Terminated
not on file
Reason
Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure
Root cause
Under Investigation by firm
Action
AngioDynamics issued Urgent Medical Device Recall Letter on March 2, 2023 via Federal Express. Letter states reason for recall health risk and action to take: IMMEDIATELY Stop using the product subject to recall. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location. Segregate this product in a secure location for return to AngioDynamics, Inc. Forward a copy of this recall notification to all sites to which you have distributed affected product. 2. ComplFx Reply Verification Tracking Form: Attn: NK Probe 5-Pack Recall Coordinator Fax number 1-855-273-0519 3. Package and Return the Recalled Product. Package any product that is being returned in an appropriate shipping box. Write the RMA number on the RMA/Address label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. Seal the box and return to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: NK Probe 5-Pack Recall Coordinator
Quantity
10 boxes( 5 probes/box)
Distribution
US Nationwide Distribution: CA,CO, FL, IA, NY, OH
Lot, serial or model codes
UDI-DI: 15051684029643 Lot Number: 5762110
510(k) numbers
["K183385"]
PMA numbers
not on file
Product code
OAB
Firm FEI
1319211
Firm city
Queensbury
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register