Skip to the content

FDA Medical Device Recalls (openFDA)

Arthrex, Inc.: Tibial Opening Wedge Osteotomy Plate, Titanium, 15.0 mm - Product Usage: use in treatment of non-union, mal-union, and fractures of the proximal tibia and distal femur.

Recall number
Z-1380-2021
Recalling firm
Arthrex, Inc.
Product
Tibial Opening Wedge Osteotomy Plate, Titanium, 15.0 mm - Product Usage: use in treatment of non-union, mal-union, and fractures of the proximal tibia and distal femur.
Status
Terminated
Initiated
2021-03-11
Posted
not on file
Terminated
2024-10-29
Reason
It was discovered that five AR-13200ST-15.0 parts were processed as AR-13200T-15.0 parts.
Root cause
Under Investigation by firm
Action
A Recall notification letter titled, "URGENT: RECALL NOTIFCIATION" was sent to consignees on 03/11/2021 via email. The email sent to the consignees requested that the parts be returned. Additionally, the firm is currently awaiting all distributed units to be returned. The recalling firm requested that consignees immediately discontinue use, sale, and distribution of the product. Instructions were provided for Arthrex Agencies (North America), Arthrex Subsidiaries (LATAM and APAC), direct customers (EMEA), and Arthrex Subsidiaries (EMEA) as follows: Arthrex Agencies (North America): Please contact Arthrex Product Surveillance at 866-267-9138 or complaints@arthrex.com. The consignee is not required to notify patients of the product recall. Arthrex Subsidiaries (LATAM and APAC): Please contact Arthrex Product Surveillance at 866-267-9138 or complaints@arthrex.com. The consignee is not required to notify patients of the product recall. Direct customers (EMEA): Please contact Arthrex Product Surveillance at +49 89 90 90 05 5240 or complaints@arthrex.com. The consignee is not required to notify patients of the product recall. Arthrex Subsidiaries (EMEA): Please contact Arthrex Product Surveillance at +49 89 90 90 05 5240 or complaints@arthrex.com. The consignee is not required to notify patients of the product recall.
Quantity
32 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of GA, LA, CA, FL, OR, TX and the countries of India, Russia, Kingdom of Bahrain, Italy, Canada, Sweden, Poland, Australia, New Zealand, and Mexico.
Lot, serial or model codes
Part number: AR-13200T-15.0; Lot code:12135408
510(k) numbers
["K032187"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1220246
Firm city
Naples
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register