Skip to the content

FDA Medical Device Recalls (openFDA)

Windstone Medical Packaging, Inc.: Upper Extremity Pack, part number AMS4670 Upper Extremity Pack, part number AMS4670(A

Recall number
Z-1378-2017
Recalling firm
Windstone Medical Packaging, Inc.
Product
Upper Extremity Pack, part number AMS4670 Upper Extremity Pack, part number AMS4670(A
Status
Terminated
Initiated
2017-01-11
Posted
2017-02-28
Terminated
2017-05-03
Reason
The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.
Root cause
Nonconforming Material/Component
Action
AMS notified their consignees by Email, Fax, and Phone on 11/11/2017. AMS sent the notification letter, dated January 11, 2017, and provided yellow stickers "Recall Notice" to customers. They are advised to over-label the affected AMS kits at their facilities. Customers were instructed to place sticker in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date on a kit. Customers with questions can call Vicki Davis at 321-527-7714 or send email to vdavis@alignedmedicalsolutions.com.
Quantity
1787 kits
Distribution
Distributed in AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NV, OH, PA, TX, UT, VA, and WA.
Lot, serial or model codes
Lot number and expiration date  100052, 9/16/2018 84369, 6/20/2017 84721, 6/26/2017 84951, 6/21/2017 85377, 7/26/2017 86594, 10/8/2017 86595, 10/11/2017 87312, 10/24/2017 87990, 10/6/2017 88277, 10/26/2017 88668, 11/1/2017 89203, 12/16/2017 89605, 12/17/2017 94907, 12/28/2017 94968, 12/5/2017 95303, 2/27/2018 95476, 2/20/2018 96326, 2/1/2018 96619, 2/10/2018 96788, 2/1/2018 96950, 1/5/2018 97302, 3/20/2018 97596, 4/28/2018 97819, 3/15/2018 98048, 3/24/2018 98529, 5/30/2018 98710, 6/10/2018 98945, 5/24/2018 98946, 6/7/2018
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSY
Firm FEI
1000125955
Firm city
Billings
Firm state
MT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register