Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: GE Healthcare Discovery MR750 3.0T The GE Signa MR750 System is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The GE Signa MR750 System is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and organs of the entire body, including, but not limited to, head, neck, TMVJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the GE Signa MR750 System reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images when interpreted by a trained physician yield information that may assist in diagnosis. The indications for use for the 3.0T GE Signa MR750 System are similar to those for the Signa HDx Magnetic Resonance System and the Siemens MAGNETOM Verio MR System.

Recall number
Z-1375-2011
Recalling firm
GE Healthcare, LLC
Product
GE Healthcare Discovery MR750 3.0T The GE Signa MR750 System is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The GE Signa MR750 System is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and organs of the entire body, including, but not limited to, head, neck, TMVJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the GE Signa MR750 System reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images when interpreted by a trained physician yield information that may assist in diagnosis. The indications for use for the 3.0T GE Signa MR750 System are similar to those for the Signa HDx Magnetic Resonance System and the Siemens MAGNETOM Verio MR System.
Status
Terminated
Initiated
2010-11-18
Posted
2011-02-18
Terminated
2012-02-04
Reason
GE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may impact service or maintenance personnels safety. The gradient cooling manifolds used on the affected systems may contain electrically conductive hoses, which become live when the g
Root cause
Nonconforming Material/Component
Action
GE Healthcare sent an "Urgent Medical Device Correction" letter dated November 16, 2011, to all affected customers. The letter was addressed to Hospital Administrator / Risk Manger, Radiology Department Manager and Maintenance / Biomedical Engineering Department. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product correction and Contact Information. Customers were instructed to follow appropriate lockout/tagout (LOTO) procedures. Customers were instructed to contact GE Healthcare engineering at 843-687-7213 or 843-697-3009 for instructions on how to proceed. For questions concerning the notiifcaton customers were instructed to contact their local Support Engineer.
Quantity
25
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, LA, MD, MA, MI, MO, NV, NJ, NY, NC, OR, PA, SC, TN, TX, WA, WI. and the countries of AUSTRALIA, BRAZIL, CANADA, EHYPT, FINLAND, FRANCE, GERMANY, GREECE, INDIA, ITALY, JAPAN, KUWAIT, QATAR, REPUBLIC OF KOREA, SWITZERLAND, TAIWAN, and the UNITED KINGDOM.UNITED KINGDOM, TAIWAN, SWITZERLAND, REPUBLIC OF KOREA, QATAR, KUWAIT, JAPAN, ITALY, INDIA, GREECE, GERMANY, FRANCE, FINLAND, EGYPT, CANADA, BRAZIL, AUSTRALIA.
Lot, serial or model codes
00000XRMBB0092 00000XRMBB0093 00000XRMBB0098 00000XRMBB0102 00000XRMBB0103 00000XRMBB0106 00000XRMBB0108 00000XRMBB0113 00000XRMBB0115 00000XRMBB0117 00000XRMBB0128 00000XRMBB0134 00000XRMBB0112 00000XRMBB0107 00000XRMBB0076 00000XRMBB0061 00000XRMBB0116 00000XRMBB0077 00000XRMBB0097 00000XRMBB0062 00000XRMBB0130 00000XRMBB0118 00000XRMBB0124 00000XRMBB0095 00000XRMBB0105
510(k) numbers
["K081028"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register