Skip to the content

FDA Medical Device Recalls (openFDA)

Ventana Medical Systems Inc: Ventana HE 600 System, automated slide preparer, for use in laboratories.

Recall number
Z-1374-2022
Recalling firm
Ventana Medical Systems Inc
Product
Ventana HE 600 System, automated slide preparer, for use in laboratories.
Status
Open, Classified
Initiated
2022-06-01
Posted
2022-07-12
Terminated
not on file
Reason
There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
Root cause
Under Investigation by firm
Action
On June 6, 2022 Ventana Medical Systems, Inc. (via Roche Diagnostics Corp.) issues an "Urgent Medical Device Correction" notification to affected consignees. Ventana Medical ask consignees to take the following actions: 1. Notify all operators of this potential hazard and forward this UMDC as appropriate. If your facility has distributed the affected device to another site, please ensure this UMDC is provided to that site. 2. Do not leave the instrument unattended when the power is on. Turn off the power to the instrument when it is unattended per the instructions in this UMDC. 3. Do not place the system in stand-by mode overnight when unattended until further mitigation actions have been implemented. 4. If fire is detected, follow your local emergency guidelines. 5. If a leak, smoke, or odor is detected in your VENTANA HE 600 instrument, power off the instrument immediately and call the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155. 6. Complete the enclosed faxback form (TP-01573) and return it according to the instructions on the form. 7. File this Urgent Medical Device Correction (UMDC) for future reference. 8. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Events Reporting Program: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or call FDA 1-800-FDA-1088.
Quantity
763 instruments
Distribution
US: AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD MI MN MO MS MT NC NE NJ NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: Canada, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Poland, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Argentina, Australia, Brazil, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Israel, Japan, Kuwait, Lebanon, Malaysia, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates (UAE), Vietnam
Lot, serial or model codes
All Lots; UDI/DI: 04015630976010
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
1000125183
Firm city
Oro Valley
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register