AMO Manufacturing USA, LLC: AMO brand WaveScan WaveFront System, OPM software versions 3.67 and 3.67, Part Number: 0070-1531, Catalog Numbers: 0070-1664,0070-1649,0070-1654: Version 3.67 0070-1721: Version 3.671 0070-1662: Version 3.67, US version; 0070-1668; IVersion 3.67 International version Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The WaveScan WaveFront System is a class III accessory device used in conjunction with the VISX STAR S4 and STAR S4 IRTM Excimer Laser Systems, which are indicated for wavefront-guided laser assisted in situ keratomileusis (LASIK) eye surgery in certain patient populations. The WaveScan WaveFront System is a diagnostic instrument intended for the automated measurement, analysis, and recording of refractive errors of the eye. The WaveScan System software uses the diagnostic data of the eye's refractive errors and wavefront aberrations to calculate the target treatment shape and then generates the commands for the VISX STAR laser to create the target shape on the cornea.

Recall number
Z-1374-2011
Recalling firm
AMO Manufacturing USA, LLC
Product
AMO brand WaveScan WaveFront System, OPM software versions 3.67 and 3.67, Part Number: 0070-1531, Catalog Numbers: 0070-1664,0070-1649,0070-1654: Version 3.67 0070-1721: Version 3.671 0070-1662: Version 3.67, US version; 0070-1668; IVersion 3.67 International version Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The WaveScan WaveFront System is a class III accessory device used in conjunction with the VISX STAR S4 and STAR S4 IRTM Excimer Laser Systems, which are indicated for wavefront-guided laser assisted in situ keratomileusis (LASIK) eye surgery in certain patient populations. The WaveScan WaveFront System is a diagnostic instrument intended for the automated measurement, analysis, and recording of refractive errors of the eye. The WaveScan System software uses the diagnostic data of the eye's refractive errors and wavefront aberrations to calculate the target treatment shape and then generates the commands for the VISX STAR laser to create the target shape on the cornea.
Status
Terminated
Initiated
2010-07-13
Posted
2011-02-18
Terminated
2012-06-22
Reason
The failure to comply with Minimum System Requirements, increased system errors could result, and a miscreated treatment table (incorrect LASIK refractive treatment) could result.
Root cause
Software design
Action
Abbott Medical Optics, Inc. sent an Urgent Medical Device Field Correction letter to all affected customers starting July 15, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to take the following actions: Surgeons were instructed to complete the facsimile attachment to acknowledge receipt of the notification letter and fax to 714-247-4530. For those using the affected product a 3.68 OPM disc must be requested by e-mailing a request to OPM@pacful.com. For any questions please contact AMO Technical Service Representatiave at 1-877-266-4543.
Quantity
not on file
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Great Britian, Greece, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Malaysia, Mexico, Netherland, New Zealand, Peru, Philippines, Portugal, Russia, Russia Federation, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, ThailandChina, India, South Korea, Singapore, Taiwan, Thailand, Turkey, United Arab Emirates, Venezuela, and Vietnam.
Lot, serial or model codes
All units of these Catalog Numbers. Part Number: 0070-1531, Catalog Numbers:  0070-1664,0070-1649,0070-1654: Version 3.67 0070-1721: Version 3.671  0070-1662: Version 3.67, US version; 0070-1668; IVersion 3.67 nternational version
510(k) numbers
not on file
PMA numbers
["P930016S016"]
Product code
LZS
Firm FEI
3006695864
Firm city
Milpitas
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register