Boston Scientific Corporation: Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN referenced on the Pouch Label) . Boston Scientific PT Graphix Magnet Guidewire with ICE Hydrophilic Coating is a steerable guidewire available with a nominal diameter of 0.014 in (0.37 mm) and in nominal lengths of 182 cm. PT Graphix Magnet is available with a shapeable Straight Tip or a preformed "J" Tip. The PT Graphix Magnet Guidewire features a 10 or 38 cm polymer sleeve, coated with ICE Hydrophilic Coating, jacketing the distal tapered core wire. The proximal section of the PT Graphix Magnet Guidewire is coated with a fluorinated polymer.

Recall number
Z-1370-2016
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN referenced on the Pouch Label) . Boston Scientific PT Graphix Magnet Guidewire with ICE Hydrophilic Coating is a steerable guidewire available with a nominal diameter of 0.014 in (0.37 mm) and in nominal lengths of 182 cm. PT Graphix Magnet is available with a shapeable Straight Tip or a preformed "J" Tip. The PT Graphix Magnet Guidewire features a 10 or 38 cm polymer sleeve, coated with ICE Hydrophilic Coating, jacketing the distal tapered core wire. The proximal section of the PT Graphix Magnet Guidewire is coated with a fluorinated polymer.
Status
Terminated
Initiated
2016-03-28
Posted
not on file
Terminated
2016-11-07
Reason
Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip model, the guidewires are straight and do not have the pre-formed J-tip. The most severe injury expected to occur is a limited procedure delay caused by the need to replace the guidewire with one that has
Root cause
Process control
Action
Boston Scientific sent an "Urgent Medical Device Recall" letter dated March 28, 2016. The letter was addressed to Cath Lab Materials Manager / Field Action Contact. The letter described the problem and the product affected by the recall. Advised consignees to cease immediately further distribution or use of the product and to notify their customers. They also requested consignees to complete and return "The Reply Verification Tracking Form" For questions contact your local Boston scientific sales Representative of the Boston Scientific Filed Action Center..
Quantity
47 units
Distribution
US nationwide in the states of ME, LA, and TN
Lot, serial or model codes
Lot: 17971870
510(k) numbers
K143587
PMA numbers
not on file
Product code
DQX
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source