Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics Triathlon Femoral Distal Augment- Left; 1 single unit per package Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430 USA Stryker France, ZAC Satolas Green Pusignan, Av de Satoles Green 69881 Meyzieu, Cedex, France This device is used when there is a deficiency in the bone present during femoral preparation these devices are used to "fill the gap" between the implant and the bone

Recall number
Z-1370-2013
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics Triathlon Femoral Distal Augment- Left; 1 single unit per package Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430 USA Stryker France, ZAC Satolas Green Pusignan, Av de Satoles Green 69881 Meyzieu, Cedex, France This device is used when there is a deficiency in the bone present during femoral preparation these devices are used to "fill the gap" between the implant and the bone
Status
Terminated
Initiated
2013-03-25
Posted
2013-05-23
Terminated
2015-05-06
Reason
Stryker Orthopaedics received a report that a label on the external packaging (box) of a Triathalon Femoral Distal Augment 10mm- Size 5 Left depicted a glyph of an acetabular shell rather than a femoral distal augment. In addition the label was missing a reference to the "10mm" augment thickness which per the label print is required to appear below the size reference.
Root cause
Software design
Action
The firm, Stryker Orthopaedics, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 25, 2013 and Product Accountability Forms via Fed Ex to its Consignees/ customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete and return the attached Notification Acknowledgment Form within 5 days via fax to 855-251-3635; contact the hospitals in your territory that have the affected product to arrange return of product; and return affected product to the attention of Regulatory Compliance Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430, attach flourescent orange PRODUCT REMEDIATION sticker to your return indicating the PR #RA2013-011 and mark outer box with the words "Product Recall". If you have any questions, feel free to contact Manager, Divisional Regulatory Compliance at (201) 972-2100.
Quantity
11 units
Distribution
Nationwide distribution: USA including states of: HI, IN, MS, NC, NJ, OK, TX, and WI.
Lot, serial or model codes
Catalog Number 5541-A-501 Lot Codes HALY, HAOO, HANG
510(k) numbers
K070095
PMA numbers
not on file
Product code
JWH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register