FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370
- Recall number
- Z-1362-2018
- Recalling firm
- Beckman Coulter Inc.
- Product
- BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370
- Status
- Terminated
- Initiated
- 2018-03-02
- Posted
- not on file
- Terminated
- 2023-07-06
- Reason
- Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.
- Root cause
- Use error
- Action
- The customers were contacted initially by the Urgent Medical Device Recall letter on 03/028/2018 delivered by email and US Mail. Beckman Coulter will schedule a sites visit to verify the reporting units of measurement for casts for the laboratory within 90 days of this recall initiation. The use of Casts/HPF is not a common setting worldwide.
- Quantity
- 5247 units
- Distribution
- Worldwide
- Lot, serial or model codes
- All
- 510(k) numbers
- ["K022774"]
- PMA numbers
- not on file
- Product code
- KQO
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28