Skip to the content

FDA Medical Device Recalls (openFDA)

Orthovita, Inc., dBA Stryker Orthobiologics.: Vitagel; indicated in surgical procedures (other than in neurosurgical and ophthalmic) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.

Recall number
Z-1353-2015
Recalling firm
Orthovita, Inc., dBA Stryker Orthobiologics.
Product
Vitagel; indicated in surgical procedures (other than in neurosurgical and ophthalmic) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.
Status
Terminated
Initiated
2015-02-26
Posted
2015-03-30
Terminated
2015-07-28
Reason
Several shipments of Vitagel product (a surgical hemostat) required to be stored at 2 -8 degrees Celsius (C) were not delivered to customers within the validated delivery time.
Root cause
Storage
Action
Orthovita, Inc., dBA Stryker Orthobiiologics made phone notifications in February 2015 to all affected customers. Customers were told that the Vitagel they ordered on x date has an internal procedural mandate that says they must deliver in 35 hours and it has now passed that window. As a precaution until we can confirm that Vitagel delivered after shipping for longer is safe for use we are asking that they quarantine the order in question in the event the delivery occurs. For questions regarding this recall call 610-640-1775.
Quantity
274
Distribution
Nationwide Distribution including NY PA WV MI IL MA TX AR KS FL WI CA MO LA VA IA OH and NC.
Lot, serial or model codes
Product Numbers: 2113-0000, 2113-0202 , 2113-0205; Lot Numbers: A1407061, A1406053, A1410055, A1501033, A1407037, A1412010, A1412076; Expiry Dates: 02/28/2017, 08/31/2016, 03/31/2017, 10/31/2016, 10/31/2016, 01/31/2017, 09/30/2016
510(k) numbers
not on file
PMA numbers
["P050044"]
Product code
LMH
Firm FEI
1000307073
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register