Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America, LLC: Ambulance Parameter Extension (APE) kit M4758A, an accessory to the Philips HeartStart MRx Monitor/Defibrillator - Product Usage: is intended to be used with the M3536A MRx in EMS environments.

Recall number
Z-1342-2020
Recalling firm
Philips North America, LLC
Product
Ambulance Parameter Extension (APE) kit M4758A, an accessory to the Philips HeartStart MRx Monitor/Defibrillator - Product Usage: is intended to be used with the M3536A MRx in EMS environments.
Status
Terminated
Initiated
2020-01-14
Posted
not on file
Terminated
2021-02-09
Reason
M4758A Ambulance Parameter Extension (APE) kit may cause ECG "Leads Off" condition in hot, humid environments
Root cause
Process design
Action
Philips issued notification letter dated January 2020 to Philips Markets Organizations responsible for distributing the letters outside of the U.S Letter stated problem, health risk and is asking customers to follow the Action to be Taken by Customer/User section of the CIL: Philips recommends that the APE kit be removed from use and disposed of. Patients can of course be monitored and treated in ambulances using the MRx Monitor/Defibrillator without the APE kit. To acknowledge receipt of this notification, please complete and fax the Customer Reply Form back to Philips at your earliest convenience.
Quantity
43 units
Distribution
International distribution in the countries of Austria, Germany, Netherlands, Norway, Poland, Switzerland
Lot, serial or model codes
All
510(k) numbers
K063375
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register