Leica Biosystems Richmond Inc.: Surgipath Cyto Jar Fixative; a 120 mL clear, leakproof container prefilled with 45 mL of preservative containing polyethylene glycol for the convenient collection, fixation and preservation of all cytologic specimens, 50 jars per case; for in vitro diagnostic use only; Leica Biosystems, 5205 Rt. 12, Richmond, IL 60071; Made in USA The Cyto Jar is for the convenient collection, fixation and preservation of all cytologic specimens. The 120 mL clear, leakproof container is prefilled with 45 mL of preservative containing polyethylene glycol. It can be used for rinsing aspiration needles or bronchial brushes and for the direct collection of sputum specimens.

Recall number
Z-1331-2012
Recalling firm
Leica Biosystems Richmond Inc.
Product
Surgipath Cyto Jar Fixative; a 120 mL clear, leakproof container prefilled with 45 mL of preservative containing polyethylene glycol for the convenient collection, fixation and preservation of all cytologic specimens, 50 jars per case; for in vitro diagnostic use only; Leica Biosystems, 5205 Rt. 12, Richmond, IL 60071; Made in USA The Cyto Jar is for the convenient collection, fixation and preservation of all cytologic specimens. The 120 mL clear, leakproof container is prefilled with 45 mL of preservative containing polyethylene glycol. It can be used for rinsing aspiration needles or bronchial brushes and for the direct collection of sputum specimens.
Status
Terminated
Initiated
2012-02-20
Posted
2012-03-29
Terminated
2012-09-20
Reason
Product was distributed with the incorrect lot number and expiration date. The lot number and expiration date should have been listed at 073013 and instead was noted as 073014 which is beyond the maximum expiration dating of 18 months for this product.
Root cause
Process design
Action
The firm, Leica Microsystems, Inc., sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 17, 2012, via FedEx on February 20, 2012 to all affected customers. The letter describes the product, problem, and actions to be taken. A second letter was sent to the customers on February 29, 2012, clarifying that the incorrect lot number and expiration date was 073014, and should have been 073013. The customers were instructed to discontinue use of the product; return the product and complete and return the enclosed Acknowledgement Form via fax to 815-678-2216 within one week of receiving the notification. If you have any questions, contact the Regulatory Affairs Manager at toll free (800) 225-8867 or email: richmond@leica-microsystems.com.
Quantity
11 cases
Distribution
Nationwide distribution: USA including states of: Michigan, Missouri and New York.
Lot, serial or model codes
item #3801825, reorder #01825, lot #073013
510(k) numbers
not on file
PMA numbers
not on file
Product code
LEA
Firm FEI
1419341
Firm city
Richmond
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register