Iradimed Corporation: Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50

Recall number
Z-1310-2023
Recalling firm
Iradimed Corporation
Product
Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,
Status
Open, Classified
Initiated
2023-02-24
Posted
2023-03-30
Terminated
not on file
Reason
Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.
Root cause
Process control
Action
Iradimed Corporation issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/24/2023 by certified mail. The notice explained the issue and the risk and provided updated instructions for use to mitigate the potential risk of inlet occlusion. For further information concerning this issue or have any questions, please contact Iradimed support via email at SyringeAdapterRecall@Iradimed.com or via telephone at 407-677-8022.
Quantity
3285 boxes of 50 devices
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates.
Lot, serial or model codes
UDI di 10856038001048, Lot Numbers: 21K50N, exp. 11/30/2023; 22A03N, exp. 01/31/2024; 22A59N, exp. 2/29/2024; 22B54N, exp. 2/29/2024; 22D15N, exp. 4/30/2024; 22D28N, exp. 4/30/2024; 22D58N, exp. 4/30/2024; 22I17N, exp. 9/30/2024; 031851, exp. 9/30/2024; 22J46N, exp. 10/31/2024; 032198, exp. 11/30/2024; 22L02N,exp. 12/31/2024; 032822, exp. 1/31/2025
510(k) numbers
["K050301","K090087","K143369"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3005053560
Firm city
Winter Springs
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register