Meridian Bioscience Inc: The ImmunoCard Mycoplasma EIA detects the presence of IgM to M. pneumoniae in serum - Product Usage: intended to aid in the diagnosis of recent Mycoplasma pneumoniae infection.

Recall number
Z-1299-2021
Recalling firm
Meridian Bioscience Inc
Product
The ImmunoCard Mycoplasma EIA detects the presence of IgM to M. pneumoniae in serum - Product Usage: intended to aid in the diagnosis of recent Mycoplasma pneumoniae infection.
Status
Terminated
Initiated
2021-02-08
Posted
not on file
Terminated
2021-10-26
Reason
Use of the kits after incorrect storage could lead to a potential false-negative or invalid result. In the event of a false-negative, the incorrect assumption would be made as to the cause of patient symptoms. The appropriate treatment may not be administered, and symptoms may progress. Additional evaluations will be completed and documented within the associated Health Hazard Evaluation.
Root cause
Storage
Action
On 02/19/2021, Meridian issued an Urgent Medical Device Correction notice to customers via letter for ImmunoCard Mycoplasma kits from a specified lot. Upon receipt at the Cardinal Health Charlotte distribution center, the kits were identified to have a transit time of 115 hours with a maximum temperature of 55 F (approximately 12 C). This is beyond our transit time requirement of 72 hours and storage at 2-8 C. Cardinal placed the kits into a quarantine location, however 8 kits were inadvertently shipped to your facility.
Quantity
8 kits
Distribution
US Nationwide distribution in the states of OH, SC.
Lot, serial or model codes
Catalog Number: 709030, Lot Number: 709030M114, Device Identifier: 00840733101762.
510(k) numbers
["K934550"]
PMA numbers
not on file
Product code
LJZ
Firm FEI
1524213
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register