Skip to the content

FDA Medical Device Recalls (openFDA)

Corin Ltd: Corin BIOLOX DELTA MOD HEAD 36mm EX LONG +8mm12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity Biolox Delta Ceramic Head (Generic Name: Hip Prosthesis). The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell; acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW; and ceramic and CoCrMo modular heads. The ceramic and CoCrMo modular heads are compatible with Corin titanium femoral stems. The Trinity Biolox Ceramic heads are packaged in a double peel pouch configuration.

Recall number
Z-1291-2021
Recalling firm
Corin Ltd
Product
Corin BIOLOX DELTA MOD HEAD 36mm EX LONG +8mm12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity Biolox Delta Ceramic Head (Generic Name: Hip Prosthesis). The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell; acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW; and ceramic and CoCrMo modular heads. The ceramic and CoCrMo modular heads are compatible with Corin titanium femoral stems. The Trinity Biolox Ceramic heads are packaged in a double peel pouch configuration.
Status
Terminated
Initiated
2021-01-25
Posted
not on file
Terminated
2022-11-23
Reason
The size indicated on the labeling on the outer packaging, on the inner packaging, and on the patient stickers inside the packaging is incorrect: the Trinity Biolox Delta Ceramix Head 32XL is labeled as 36XL and vice-versa.
Root cause
Labeling mix-ups
Action
The firm contacted their consignee by email on 01/25/2021. The distributor was instructed to identify the affected product in its inventory and return it to Corin USA in order to contain the part. A recall letter will be sent and acknowledgement of receipt will be requested.
Quantity
23 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Colombia, Poland, Germany, Australia.
Lot, serial or model codes
Model:104.3615/Lot number:464665
510(k) numbers
["K130343"]
PMA numbers
not on file
Product code
MEH
Firm FEI
3002806725
Firm city
Cirencester
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register