Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2

Recall number
Z-1274-2021
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2
Status
Terminated
Initiated
2021-02-08
Posted
2021-03-18
Terminated
2025-02-07
Reason
Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correction for ADVIA Centaur4P Herpes-2 lgG (HSV2) Onboard Stability and Calibration Frequency (Interval) Failures on the ADVIA Centaur XP System and ADVIA Centaur XPT System.
Root cause
Under Investigation by firm
Action
In February 2021, Siemens will issue an Urgent Medical Device Correction notice to customers via letter stating that Siemens Healthcare Diagnostics Inc. has confirmed that the ADVIA Centaur HSV2 assay does not meet the 60 day onboard stability or the 28 day calibration frequency (interval) as stated in the ADVIA Centaur XP/XPT Herpes-2 IgG (HSV2).
Quantity
390 kits total (Domestic 366 kits, Foreign 24 kits)
Distribution
Domestic: AL, CA, FL, IL, MD, NJ, NY, TN, TX. Foreign: Nepal, Chile, Georgia, Spain.
Lot, serial or model codes
***Updated as of 3/25/2021*** Lot #: B33255, B33269, B33413. UDI #:  (01)00630414016726(10)B33255(17)20210625,  (01)00630414016726(10)B33269(17)20210629, (01)00630414016726(10)B33413(17)20210819. Updated as of 3/25/2021
510(k) numbers
["K181334"]
PMA numbers
not on file
Product code
MYF
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register