GE Healthcare, LLC: NM/CT 870 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce

Recall number
Z-1265-2019
Recalling firm
GE Healthcare, LLC
Product
NM/CT 870 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce
Status
Terminated
Initiated
2019-03-20
Posted
not on file
Terminated
2020-10-05
Reason
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Root cause
Device Design
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
27 units
Distribution
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Lot, serial or model codes
Mfg. Lot or Serial # System ID Model, Catalogue or Code # 870Y64037 SHIPPED (GON 4566933) H3100AS 870Y64040 SHIPPED (GON 4751375) H3100AS 870X64011 216844DR870 H3100AS 870X64024 802775D870 H3100AS 870X64002 206SCCANM H3100AS 870X64032 262741NUC H3100AS 870X64001 030020NU06 H3100AS 870Y64034 514345BD870 H3100AS 870Y64035 M54606132 H3100AS 870X64012 X634445001 H3100AS 870X64007 V2728113 H3100AS 870X64009 M4016255 H3100AS 870X64026 M40230109 H3100AS 870X64017 M40240121 H3100Q 870X64018 M40240122 H3100Q 870X64016 499221NU01 H3100AS 870X64022 A5173220 H3100AS NT3DG1800003BH NS0303 5376204-70-60-1 NT3DG1800002BH NS0302 5376204-70-60-1 870X64025 EMM0530 H3100AS 870Y64038 MST01300 H3100AS 870X64029 PL1031NM01 H3100AS 870X64027 PL0398NM01 H3100AS 870X64030 PL2190NM01 H3100AS 870X64028 PL0372NM01 H3100AS 870X64013 NP473631 H3100AS 870X64008 00669NUC03 H3100AS
510(k) numbers
K093514
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register